background image

 - 6 - 

D. Data Hold 

While measuring, press the 

HOLD 

button (3) to freeze the displayed 

value.  The LCD will also display the word: "HOLD." 

 

Press the 

HOLD 

button (3) again to exit

.

 

 
E. Maximum / Minimum 

To record the maximum and minimum readings:  

Press the 

REC 

button (4) once. "REC" appears on the LCD. 

Press the 

REC 

button (4) again. "MAX", "REC" and the maximum 

measurement appear on the LCD.   

Press the 

REC 

button (4) again.  "REC Min" and the minimum meas-

urement appear on the LCD.   

To exit the this function, press and hold the 

REC

 button (4) for at least 

2 seconds, until the display reverts to the current reading. 

 

F. Data Logger 

The data logger function can save up to 16,000 data points along with the 
time (hour-min-sec) and date (year-month-day).  There are 16,000 data 
points in memory when the display indicates "FULL."   
 

Data Logger Procedures

To momentarily display the set sampling time, press the 

LOGGER 

button 

(8) once.  To change the sampling time, see "Sample Time Setting" (pg 7).   

 

Press the 

REC 

button (4) once, "REC" will be displayed.  Depending on 

your set sampling time, continue to step "a" or "b" (below). 

 

a.  

Auto Data Logger (sampling time can be set for 2 seconds to 
8 hours, 59 minutes, 59 seconds):

  Press the 

LOGGER 

button (8) 

once, "REC DATA" is displayed and "Recording. . . ." flashes at each 
set interval. 

      Press the 

LOGGER 

button (8) once to exit.  "REC" is still displayed, but 

"DATA" is no longer shown.  To resume data logging, simply press the 

LOGGER 

button (8) again.  To exit the "REC" function, press and hold 

the 

REC 

button (4) for at least 2 seconds. 

 

 

b.

   Manual Data Logger (sampling time MUST be set to "0" zero): 

 

Press the 

LOGGER 

button (8) once to save each data point.  

    To exit the this function, press and hold the 

REC 

button (4) for at  

least 2 seconds. 

Summary of Contents for 800022

Page 1: ...Datalogging Dew Point Thermohygrometer 800022 SPER SCIENTIFIC LTD...

Page 2: ...Data Logger 6 4 ADVANCED ADJUSTMENTS 7 A Check Remaining Available of Data Points 7 B Clear Memory 7 C Date Time Setting 7 D Sample Time Setting 7 E Auto Power Off Default Setting 7 F Temp Unit Defau...

Page 3: ...temperature readings and also reads type K or J thermocouple probes Features min max calendar clock selectable auto power off and hold function This well made unit is extremely accurate has fast respo...

Page 4: ...4 2 PANEL DESCRIPTION...

Page 5: ...t the PROBE PLUG 19 into the PROBE INPUT SOCKET 17 Slide the PROBE LOCK SWITCH 18 to the LOCK position Turn on the meter by pressing the POWER button 2 The readings are displayed as RH and C or F B De...

Page 6: ...ts in memory when the display indicates FULL Data Logger Procedures To momentarily display the set sampling time press the LOGGER button 8 once To change the sampling time see Sample Time Setting pg 7...

Page 7: ...ate time set screen Use the UP 5 DOWN 6 and ENTER 4 buttons to select the year month date hours minutes and seconds D Sample Time Setting Press the SET button 8 as needed to reach the sample time scre...

Page 8: ...Press the SEND button 7 for at least 2 seconds until the bottom right dis play shows Transmit mode Release the button The following screens flash alternately The meter can record up to 16 000 data poi...

Page 9: ...stor The 16 digits data stream will be displayed in the following format D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0 Each digit indicates the following status D0 End Word 0D D1 D8 Display re...

Page 10: ...or small object into the SYSTEM RESET 13 while turn ing on the meter 9 OPTIONAL ACCESSORIES 840097 AC to DC 9V Adapter 840090 Water Resistant Instrument Pouch 840092 Bench Top Tripod 840093 Field Trip...

Page 11: ...s Type J 58 2012 F 100 1100 C Temperature Compensation Automatic for the humidity function and the type K J thermometer Type K J specification tests under the environment RF Field Strength less than 3...

Page 12: ...nt Sper Scientific product will be substituted if available This warranty does not cover probes batteries or damage resulting from accident misuse or abuse of the product In order to obtain warranty s...

Reviews: