X-ray Pre-warning:
delay time from X-ray ON and the actual turn on of HV. The Pre-warning
indicator will be on during this time. The time values are in seconds.
KV ramp rate:
time in seconds for the kV high voltage output to go from 0 to FS rated output
voltage.
mA ramp rate:
time in seconds for the mA output current to go from 0 to FS rated output
current.
Arc Count:
number of arcs before arc fault is issue and HV is turned off.
Quench time:
time that HV is held off after an arc has been detected, applies only when Arc
Counter >1, Arc Counter will reset 100X this set value (100sec max).
Max kV:
maximum kV setting for the unit must be less or equal the high voltage power supply
output ratings for kV.
Max mA:
maximum mA setting for the unit must be less or equal the high voltage power supply
output ratings for mA.
Save user parameters:
click in box “click here to send and save these values to the power
supply” all user parameter are saved in memory inside the HVPS.
Recall user parameters:
click in box “click here to request user parameters” all user
parameters are recalled from HVPS memory to GUI controller, for viewing.
Filament Parameters
Large Power Limit:
This is maximum power that can be set (product of kV multiplied mA)
when Large Filament is selected. Over power fault will occur if actual power exceeds 5% above
this set value. This value is stored in HVPS memory.
Small Power Limit:
This is maximum power that can be set (product of kV multiplied mA)
when small Filament is selected. Over power fault will occur if actual power exceeds 5% above
this set value. This value is stored in HVPS memory.
Large Filament Limit:
Maximum Current the HVPS will produce before it it goes into current
limit mode. This value must be less than or equal to 6 Amperes.
Small Filament Limit:
Maximum Current the HVPS will produce before it goes into current
limit mode. This value must be less than or equal to Maximum current specified (Amperes).
XRV SERIES
Page 30
118116-001 REV C