Input and Output Connections 
 

4.1 

The low voltage signal connections are made by a 15 

way ‘D’ connector; the pin out is shown below. 

 

Pin # 

Signal Name 

Range 

+24V input 

 

Power Ground 

 

Filament (direct program) 

 

Signal ground 

 

HV enable 

Digital input 

Not used 

 

Cathode voltage program output 

– 10V from internal pre set 

Cathode voltage program input 

– 10V input 

Cathode voltage monitor 

– 10V output 

10 

Emission current monitor 

– 10V output 

11 

Filament current monitor 

– 10V output 

12 

Emission current program input 

– 10V input 

13 

Emission current program output 

– 10V from internal pre set 

14 

Signal ground 

 

15 

Pre-set maximum filament current set value 

 

 

4.2 

Output:  

The HV output is a flying lead consisting of 2 Reynolds 18kV rated FEP insulated wire, 1.02mm overall 
diameter. Conductors 19/40AWG. Overall length 500mm. The two wires are sleeved together.  

 

Pin 

Name 

N/A 

F/Cathode 

N/A 

Filament-/Cathode 

 

 

 

 

 

Summary of Contents for XMPF10N5/24

Page 1: ... INSTALLATION INSTRUCTIONS Document Number 81234 4 Issue A Date 23 01 18 Issuing Authority I 5650 Checked Approved Tel 44 0 1798 873986 Fax 44 0 1798 872479 E mail hvsales spellmanhv co uk Broomers Park Broomers Hill Lane Pulborough W Sussex RH20 2RY ...

Page 2: ...tions unless equipment is turned off and the capacitance of both the load and power supply are grounded Allow adequate time for discharge of internal capacitance of the power supply Do not ground yourself or work under wet or damp conditions Servicing Safety Maintenance may require removing the Instrument cover with the power on Servicing should only be done by qualified personnel aware of the haz...

Page 3: ...and the conditions of this manual must be complied with to maintain safety Operating the unit in a manner not specified in this manual may impair the protection against electric shock provided by the unit The unit is contained in an earthed case the system protective earth shall be provided to the chassis The case of the unit shall be properly bonded to the main protective earth termination in the...

Page 4: ...arranty 3 2 Mechanical Installation The unit should only be used in a Pollution Degree 2 Installation Category II environment The input and output connectors are not intended for field connections and should only be connected to internal wiring in the end product The unit is intended for use as a component and no surface of the unit should be accessible in the end product 3 3 Electrical Installati...

Page 5: ...de voltage program input 0 10V input 9 Cathode voltage monitor 0 10V output 10 Emission current monitor 0 10V output 11 Filament current monitor 0 10V output 12 Emission current program input 0 10V input 13 Emission current program output 0 10V from internal pre set 14 Signal ground 15 Pre set maximum filament current set value 4 2 Output The HV output is a flying lead consisting of 2 Reynolds 18k...

Reviews: