PAGE 5 OF 5
DIMENSIONS: in.[mm]
DIMENSIONS: in.[mm]
HIGH VOLTAGE ELECTRONICS CORP.
Spellman
DANGER
HIGH
VOL
T
AGE
DANGER
HIGH
VOLTAGE
J1
0-50KV
Output
FIL I LIM ADJ
KV ADJ
mA ADJ
PH ADJ
HV ON
POWER ON
FIL RET
FIL OUT
GND
+24V
+24V @ 4A
+24V Input Return
+24V @ 4A Input Power
4 PIN
1
2
3
4
1
2
3
6
9
1K
1K
1K
10K
10K
14
5
7
8
10
11
15 PIN
X-Ray Tube
Filament
VOLTAGE
MONITOR
15 PIN
15 pin
15 pin
12
13
4
J1
14V
12V LAMP
.5W-2W
12
13
4
15
15
14V
J1
+24VDC
+24VDC
+24VDC
12
13
4
14V
15
270, 1W
+24VDC
+24VDC
+12V,
User
Provided
+24V, Power to
HV Circuits
+24V, Power to
HV Circuits
EXTERNAL
INTERLOCK
Fail Safe Interlock / HV ON Connections
Fail Safe
Interlock/HV ON:
Remove internal jumper
J1 to isolate the interlock
relay coil +
from the in24V.
Connect Pin 13 to an
ex24V source or
to Pin 12 to re-connect
in24V.
Alternate Interlock Configuration:
Fail Safe Lamp Replaced With a
270 Ohm Resistor
Alternate Interlock Configuration:
Customer Pr12V For HV ON
Relay, Relay Return Grounded
EXTERNAL
INTERLOCK
+24VDC
+24V, Power To
HV Circuits
CURRENT
MONITOR
FILAMENT
MONITOR
FILAMENT LIMIT
SETPOINT
FILAMENT PREHEAT
SETPOINT
EXTERNAL
CURRENT
CONTROL
EXTERNAL
VOLTAGE
CONTROL
V
V
M
M
M
M
M
Jumper J4-7 to J4-8
to Use Local Voltage Control
Jumper J4-10 to J4-11
to Use Local Current Control
LOCAL VOLTAGE
CONTROL
+10.00V REFERENCE
+10.00V
+10.00V
+10.00V
+10.00V
+10.00V
25K, 20T
25K, 20T
LOCAL CURRENT
CONTROL
VOLTAGE PROGRAM IN
CURRENT PROGRAM IN
FIL. P.S.
5.0V/3.5A
+
-
uX Power Supply
uX Power Supply
uX Power Supply
uX Power Supply
ANODE
X-RAY TUBE
FILAMENT
I/O
Typical uX Operating Setup
See Wiring Diagrams for Recommended Analog Interface Connections
+24V Input Connections
Filament Connections
Analog Interface
Alternate Interlock Configurations
-
+
-
+
-
+
-
+
-
+
-
+
-
+
J1
SPELLMAN HIGH VOLTAGE ELECTRONICS CORPORATION
u
X
50W/65W/75W X-RAY
GENERATOR
65KV
Corporate Headquarters
Hauppauge, New York USA
+1-631-630-3000 FAX: +1-631-435-1620
e-mail: [email protected]
www.spellmanhv.com
128108-001 REV. A
Spellman High Voltage is an ISO 9001:2008 and ISO 14001:2004 registered company
Copyright © 2014 Spellman High Voltage Electronics Corp.
Summary of Contents for uX50P50
Page 16: ...uX MANUAL 4 118151 001 REV A Figure 2 1 50kV uX DIMENSIONS ...
Page 17: ...uX MANUAL 5 118151 001 REV A Figure 2 2 65kVuX DIMENSIONS ...
Page 21: ...uX MANUAL 9 118151 001 REV A Table 3 2 DC Input Filament Connections ...
Page 22: ...uX MANUAL 10 118151 001 REV A Figure 3 3 Monitors Voltage Current Control Connections ...
Page 23: ...uX MANUAL 11 118151 001 REV A Figure 3 4 Recommended Interlock HV on Configurations ...