101501-565 Rev C
Page 51 of 58
6.6.13 Request kV Monitor
Description:
The host requests that the firmware report kV monitor.
Direction:
Host to supply
Syntax:
<STX><60><,><CSUM><ETX>
Response:
If an error is detected, the Power Supply responds with:
<STX>60,!,<ERROR>,<CSUM><ETX>
Please reference the Error Code section for a list of valid error codes.
If the message is accepted, the Power Supply responds with:
<STX>60,<ARG>,<CSUM><ETX>
Where:
<ARG> = 0 - 4095 in ASCII format
Example Response:
KV monitor is set to full scale.
<STX>60,4095,<CSUM><ETX>