101501-565 Rev C
Page 25 of 58
5.5.9 Request Model Number
Description:
The host requests that the firmware sends the unit model number.
Direction:
Host to supply
Syntax:
<STX><26><,><ETX>
Response:
If an error is detected, the Power Supply responds with:
<STX>26,!,<ERROR>,<ETX>
Please reference the Error Code section for a list of valid error codes.
If the message is accepted, the Power Supply responds with:
<STX>26,<ARG>,<ETX>
Where:
<ARG> = Consists up to fifteen ASCII characters representing the current
model number.
Example Response:
<STX>26,ST100P100X4249,<ETX>