SL SERIES MANUAL
15
118006-001 REV Y
1
2
3
4
TB1/J5
All cables should be shielded w ith the shields
High Voltage Pow er Supply.
being returned to the chassis ground of the
PS Common
Transistor
ON
Inhibits
High
Voltage
+15V
2.2k
High Voltage Pow er
Supply Internal inhibit circuitry.
The transistor should be located as close as possible
to the pow er supply
1
2
3
4
TB1/J5
All cables should be shielded w ith the shields being returned
to the chassis ground of the High Voltage Pow er Supply.
5
S1
S1 Open = Interlock Open
S1 Closed = Interlock Closed
Remove jumper
for remote
external interlock
Relay contacts are recommended for S1. The relay
should be located as close as possible to the pow er supply.
Signals are at 15Vdc, 25mA max and are only to be used
for contact closure.
External Safety
Interlock Control
Figure 3.5 -- External Inhibit Interface Circuit
Figure 3.6 -- External Interlock Interface