background image

3. Watermark:  Displays  the  soil  moisture  tension  in  kPa.

Reader  must  be  connected  to  a  WatchDog  compatible
Watermark sensor (see Watermark Sensors, p 5).

4. Temperature: This mode is valid when a temperature sen-

sor or an SMEC300 sensor is connected to the reader.

5. Raw AD: Raw electronic reading.

6. 

Calibration:  Initiate  EC  calibration  for  an  SMEC300  or
soil moisture calibration for an SM100. Although readers
with firmware v 4.4 will display  CAL EC? on this screen,
they  can  also  calibrate  the  soil  moisture  circuit  of  an
SM100.

Sample Watermark 

LCD Screen 

116.0 kP 
WatrMark 

Sample Temperature 

LCD Screen  

73°F 23C 

TEMP 

Sample Raw AD 

LCD Screen 

1082 AD 

Raw AD 

EC Calibration Initiation 
Calibration Initiation 
LCD Screen (FW v4.4) 

 LCD Screen (current) 

CAL? 

Hold SET 

CAL EC? 

Hold SET 

Summary of Contents for 6466

Page 1: ...Soil Sensor Reader PRODUCT MANUAL Item 6466 ...

Page 2: ...2 CONTENTS Basic Operation 3 Power Up Down 3 Sensor Reader Modes 3 Connecting Sensors 5 Calibrating the SMEC300 EC Sensor 6 Calibrating the SM100 7 Changing the Battery 7 Taking Readings 5 ...

Page 3: ...tronic A D reading The reader converts the raw electronic reading to measurement units using internal calibration constants The sixth mode is for calibrating the EC circuit of the SMEC300 or the soil moisture reading for an SM100 see pp 6 7 1 Spectrum Standard For the SM100 and SMEC300 the volumetric water content for a mineral soil will be dis played on the first line For the SMEC300 the EC will ...

Page 4: ...reading 6 Calibration Initiate EC calibration for an SMEC300 or soil moisture calibration for an SM100 Although readers with firmware v 4 4 will display CAL EC on this screen they can also calibrate the soil moisture circuit of an SM100 Sample Watermark LCD Screen 116 0 kP WatrMark Sample Temperature LCD Screen 73 F 23C TEMP Sample Raw AD LCD Screen 1082 AD Raw AD EC Calibration Initiation Calibra...

Page 5: ... will not affect subsequent read ings WaterScout SM100 and SMEC300 Sensors Insert the 2 5mm stereo pin into the socket at the top of the meter WaterMark Sensors If the sensor is a WatchDog compatible model of the sensor item numbers 6450WD or 6450WD20 the sensor can be plugged directly into the meter like a WaterScout sensor If the sensor terminates in 2 green wires with circular leads the sensor ...

Page 6: ... the pads with your fingers 2 Allow the calibration solution to come to room tempera ture You must calibrate with a solution with an EC of 1 41 mS cm It is important for the sensor and solution to be at the same temperature Calibration Procedure 1 Pour clean calibration solution into the container 2 Plug the sensor into the reader 3 Immerse the sensor in the calibrating solution up to sec ond mark...

Page 7: ...ingers 8 Store the sensor in plastic bag until you are ready to in stall it in the field Changing the Battery Calibrating the SM100 SM100 sensors with the 3 14 code printed on the front can be re calibrated to the factory setting The calibration proce dure is nearly identical to that used to calibrate the SMEC300 EC circuit The calibration is done with distilled water The SM100 does not have an in...

Page 8: ...gy September 18 2012 Edaphic Scientific Pty Ltd 61 2 6584 5768 info edaphic com au www edaphic com au R 3 14 Warranty This product is warranted to be free from defects in material or workmanship for one year from the date of purchase During the warranty period Spectrum will at its option either repair or replace products that prove to be defective This warranty does not cover damage due to imprope...

Reviews: