The  EC  sensor  of  the  SMEC300  sensor  has  an 
internal constant that can be  adjusted by the user 
to correct for sensor drift. To calibrate the sensor, 
you will need a container wide enough to accom-
modate  the  sensor  and  tall  enough  so  the  sensor 
can be immersed to the midway point of the sen-
sor molding. 

Sensor Preparation 

Before  starting  the  calibration  procedure,  the  following  pre-
liminary  steps  are  necessary  to  ensure  the  sensor  is  properly 
calibrated.  
1. Because residual oils on the EC sensor pads, including oil

from your fingers, will reduce the accuracy of the sensor,
clean the sensor with rubbing alcohol.  After cleaning the
sensor, do not touch the pads with your fingers.

2. Allow  the  calibration  solution  to  come  to  room  tempera-

ture.    You  must  calibrate  with  a  solution  with  an  EC  of
1.41 mS/cm. It is important for the sensor and solution to
be at the same temperature.

Calibration Procedure 

1. Pour clean calibration solution into the container.
2. Plug the sensor into the reader.
3. Immerse  the  sensor  in  the  calibrating  solution  up  to  sec-

ond mark on the molding.  This ensures the sensor’s inter-
nal thermistor is below the liquid level.  Take care that the
EC sensor pads are not touching the side of the container.

4. Set reader to CAL mode (see p.4).
5. Press  and  hold  the  SET  button  for  5  seconds.  The  LCD

will  count  down  the  time.    When  the  calibration  is  com-
plete,  the  sensor  will  transition  to  Standard  Spectrum
mode  (see  p.3).    If  the  SET  button  is  released  before  the

Calibrating the SMEC 300 

EC Sensor 

Summary of Contents for 6466

Page 1: ...Soil Sensor Reader PRODUCT MANUAL Item 6466 ...

Page 2: ...2 CONTENTS Basic Operation 3 Power Up Down 3 Sensor Reader Modes 3 Connecting Sensors 5 Calibrating the SMEC300 EC Sensor 6 Calibrating the SM100 7 Changing the Battery 7 Taking Readings 5 ...

Page 3: ...tronic A D reading The reader converts the raw electronic reading to measurement units using internal calibration constants The sixth mode is for calibrating the EC circuit of the SMEC300 or the soil moisture reading for an SM100 see pp 6 7 1 Spectrum Standard For the SM100 and SMEC300 the volumetric water content for a mineral soil will be dis played on the first line For the SMEC300 the EC will ...

Page 4: ...reading 6 Calibration Initiate EC calibration for an SMEC300 or soil moisture calibration for an SM100 Although readers with firmware v 4 4 will display CAL EC on this screen they can also calibrate the soil moisture circuit of an SM100 Sample Watermark LCD Screen 116 0 kP WatrMark Sample Temperature LCD Screen 73 F 23C TEMP Sample Raw AD LCD Screen 1082 AD Raw AD EC Calibration Initiation Calibra...

Page 5: ... will not affect subsequent read ings WaterScout SM100 and SMEC300 Sensors Insert the 2 5mm stereo pin into the socket at the top of the meter WaterMark Sensors If the sensor is a WatchDog compatible model of the sensor item numbers 6450WD or 6450WD20 the sensor can be plugged directly into the meter like a WaterScout sensor If the sensor terminates in 2 green wires with circular leads the sensor ...

Page 6: ... the pads with your fingers 2 Allow the calibration solution to come to room tempera ture You must calibrate with a solution with an EC of 1 41 mS cm It is important for the sensor and solution to be at the same temperature Calibration Procedure 1 Pour clean calibration solution into the container 2 Plug the sensor into the reader 3 Immerse the sensor in the calibrating solution up to sec ond mark...

Page 7: ...ingers 8 Store the sensor in plastic bag until you are ready to in stall it in the field Changing the Battery Calibrating the SM100 SM100 sensors with the 3 14 code printed on the front can be re calibrated to the factory setting The calibration proce dure is nearly identical to that used to calibrate the SMEC300 EC circuit The calibration is done with distilled water The SM100 does not have an in...

Page 8: ...gy September 18 2012 Edaphic Scientific Pty Ltd 61 2 6584 5768 info edaphic com au www edaphic com au R 3 14 Warranty This product is warranted to be free from defects in material or workmanship for one year from the date of purchase During the warranty period Spectrum will at its option either repair or replace products that prove to be defective This warranty does not cover damage due to imprope...

Reviews: