PROBLEM

PROBABLE CAUSE

CORRECTIVE ACTION

Low system inlet pressure.

Insuffi cient feed water 
pressure or fl ow

Clogged pre-fi lter

Closed valve ahead of RO 
System

Check feed pressure, open 
feedwater valve, check for 
restrictions.

Replace pre-fi lter cartridge.

Check for closed valves.

Low system operating 
pressure.

Pump not operating correctly

Defective pressure regulator 
valve

Repair or Replace pump.

Replace or repair valve.

Low permeate fl ow rate

Low operating pressure

Dirty or fouled membrane

Operating on cold water

See above possible causes 
for low operating pressure.

Replace it.

If possible operate with a 
feedwater temperature of 72-
80 °F (22-27 °C).

Excessive Permeate fl ow 
rate.

Defective product water 
membrane O-rings.

Damaged membrane 
(possibly due to chlorine 
damage)

Ruptured membrane

Check product water tube 
O-rings and replace if 
necessary.

Replace membrane.

Replace membrane.

Water fl owing when system is 
turned off

Inlet solenoid valve not 
closing or seating properly

Clean valve or replace.

RO System cycles ON and 
OFF.

Insuffi cient system inlet 
pressure

See above possible causes 
for low system inlet pressure.

Troubleshooting Guide:

This guide is to assist you in identifying the most common operating prob-
lems you may experience with your water purification system.

Summary of Contents for PSP-1000

Page 1: ... 894 5437 Fax us toll free 1 877 527 7873 Call us toll free 1 800 685 2783 E mail spectra spectrapure com Visit us on the web www spectrapure com 1 Operations and Maintenance Manual for PSP Series Water Purification Systems PSP 600 PSP 1000 PSP 1600 PSP 2700 Rev 0 April 21 2005 ...

Page 2: ... loop decreases membrane fouling by rapidly flushing the membrane at a high velocity greatly increasing membrane life Time delayed inlet solenoid valve operates for several seconds before pump activates protecting the pump from dry starts and pump destroying cavitation Low voltage microprocessor based control system with Auto Flush and Pump Protection Flushes membrane after each use and protects p...

Page 3: ...F to 86 F 4 5 C to 30 C Membrane Rejection greater than 98 average RO System Recovery 50 Feed flow GPM 1 0 GPM at 100 psi 77 F PSP 600 1 4 GPM at 100 psi 77 F PSP 1000 Concentrate Reject flow GPM 0 5 GPM at 100 psi 77 F PSP 600 0 7 GPM at 100 psi 77 F PSP 1000 Permeate Product flow GPM 0 5 GPM at 100 psi 77 F PSP 600 0 7 GPM at 100 psi 77 F PSP 1000 Production Rate 100 psi 1500ppm NaCl 750 GPD 77 ...

Page 4: ...ure 110 psig max Operating Design Temperature 40 F to 86 F 4 5 C to 30 C Membrane Rejection greater than 98 average RO System Recovery 50 Feed flow GPM 2 2 GPM at 100 psi 77 F Concentrate Reject flow GPM 1 1 GPM at 100 psi 77 F Permeate Product flow GPM 1 1 GPM at 100 psi 77 F Production Rate 100 psi 1500ppm NaCl 1600 GPD 77 F System Inlet port 5 8 compression Permeate Product port 1 2 compression...

Page 5: ...sure 110 psig max Operating Design Temperature 40 F to 86 F 4 5 C to 30 C Membrane Rejection greater than 98 average RO System Recovery 50 Feed flow GPM 3 8 GPM at 100 psi 77 F Concentrate Reject flow GPM 1 9 GPM at 100 psi 77 F Permeate Product flow GPM 1 9 GPM at 100 psi 77 F Production Rate 100 psi 1500ppm NaCl 2700 GPD 77 F System Inlet port 5 8 compression Permeate Product port 1 2 compressio...

Page 6: ...o the storage tank The TDS Meter indicates the purity of the product water going into the tank and aids in determining when the RO membrane may be failing The tank will fill until water reaches the high level float or pressure switch 13 This will activate the flush cycle covered later and then shut the pump and motor down The concentrate stream is split into two streams in the high pressure manifo...

Page 7: ...rive Suite 2A Tempe Arizona 85281 2941 480 894 5437 Fax us toll free 1 877 527 7873 Call us toll free 1 800 685 2783 E mail spectra spectrapure com Visit us on the web www spectrapure com 7 Schematic for All Systems Using One Open Tank ...

Page 8: ...e Suite 2A Tempe Arizona 85281 2941 480 894 5437 Fax us toll free 1 877 527 7873 Call us toll free 1 800 685 2783 E mail spectra spectrapure com Visit us on the web www spectrapure com 8 Schematic for All Systems Using Multiple Open Tanks ...

Page 9: ...Drive Suite 2A Tempe Arizona 85281 2941 480 894 5437 Fax us toll free 1 877 527 7873 Call us toll free 1 800 685 2783 E mail spectra spectrapure com Visit us on the web www spectrapure com 9 Schematic for All Systems Using Bladder Tank ...

Page 10: ...rial Drive Suite 2A Tempe Arizona 85281 2941 480 894 5437 Fax us toll free 1 877 527 7873 Call us toll free 1 800 685 2783 E mail spectra spectrapure com Visit us on the web www spectrapure com 10 PSP 600 PSP 1000 Front Side View ...

Page 11: ...dustrial Drive Suite 2A Tempe Arizona 85281 2941 480 894 5437 Fax us toll free 1 877 527 7873 Call us toll free 1 800 685 2783 E mail spectra spectrapure com Visit us on the web www spectrapure com 11 PSP 1600 Front Side View ...

Page 12: ...dustrial Drive Suite 2A Tempe Arizona 85281 2941 480 894 5437 Fax us toll free 1 877 527 7873 Call us toll free 1 800 685 2783 E mail spectra spectrapure com Visit us on the web www spectrapure com 12 PSP 2700 Front Side View ...

Page 13: ...d by the input pressure switch 5 the control board will activate the power relay which in turn will apply power to the main pump 6 and motor 7 The pump motor will boost the recirculating system pressure to 100 110 psi If at any time during a fill cycle the input pressure drops enough to de activate the input pressure switch the main pump and motor will be powered down The control board will then a...

Page 14: ...rial Drive Suite 2A Tempe Arizona 85281 2941 480 894 5437 Fax us toll free 1 877 527 7873 Call us toll free 1 800 685 2783 E mail spectra spectrapure com Visit us on the web www spectrapure com 14 All Systems Electrical Schematic ...

Page 15: ...ter quality Record all information in a system history log Annually In addition to above sanitize all filter housings with Kay 5 sanitizer one packet to 2 5 gallons of warm water or similar product Follow standard sanitizing procedure clean with mild detergent rinse with clear water fill filter with sanitizer solution and allow to air dry or reassemble Do not rinse the parts after sanitizer soluti...

Page 16: ...8 poly tubing to brass compression fittings 1 Cut the tubing end squarely and remove internal and external burrs 2 Remove compression nut from the compression fitting and slide it over the tubing 3 Slide the white plastic collar over the tubing NOTE Smaller diameters have a tapered collar while larger diameters 5 8 have dual tapered collars Tapered collars must slope towards the fitting See typica...

Page 17: ... carbon filter for at least 15 minutes after air has been bled from the tank and until the backwash is clear of black carbon fines 5 Return the carbon filter control head back to the service position by rotating the black manual backwash knob clockwise around to the service position 6 Set time of day on the control valve on the carbon filter by depressing the red button and aligning the 24 hour ge...

Page 18: ...d shaft of the pressure relief valve located beneath the high pressure manifold 17 Adjust the pressure relief valve to 100 psi by turning the brass handle Rotate clockwise when viewed from the bottom to increase the pressure and counter clockwise to decrease the pressure 18 When adjusted to the proper pressure lock the nut back in place 19 Check the product water to waste water ratio It should be ...

Page 19: ...ow permeate flow rate Low operating pressure Dirty or fouled membrane Operating on cold water See above possible causes for low operating pressure Replace it If possible operate with a feedwater temperature of 72 80 F 22 27 C Excessive Permeate flow rate Defective product water membrane O rings Damaged membrane possibly due to chlorine damage Ruptured membrane Check product water tube O rings and ...

Page 20: ...orrectly Power switch on tank water level below the low float unit not operating Float switch has cut power to machine Defective Float Thermal overload in motor No power to system Motor and or pump not operating properly Micro switch on either carbon filter or water softener not functioning properly Carbon filter or water softener control valve not in service position Low system inlet pressure Che...

Reviews: