Spectra® 20K Library Transport Tray Assembly Replacement Guide
32
Picker-Slot Calibration
1.
From the Diagnostics screen, select
Advanced
.
2.
From the Available tests list, select
Picker Slot Calibration: Long Axis
(as shown
in Figure 39 on page 30).
3.
Select
Prepare Test
.
4.
Select
Run Test
(there are no parameters for this test).
This test will calibrate the top and bottom location of each tape pack by loading the
tape in Slot 1 and Slot 15 of each data cartridge magazine.
5.
Watch to see that each first and fifteenth slot loads correctly.
The tape will be returned to the E/E port when the test is completed.
Note:
This test may take a long amount of time to complete.
6.
Check the test results.
The test may indicate that retries were made, which is normal.
Caution:
Make sure that there are no slots that
fail
to load. If a slot fails,
contact Spectra Logic Technical Support at (800) 227-4637 or at
(303) 449-0160.
7.
Return to the Inventory screen.
8.
Select
Open/Close
to open the E/E port.
9.
Remove the tape from the library.
10.
Select
Open/Close
to close the E/E port.