Troubleshooting Guide
Clinical Situation
Possible Cause
Solution
No SpO
2
label is
displayed
■
SpO
2
is not enabled at the 91343
■
Be sure transmitter DIP switch 1 and 2
are set correctly
■
SpO
2
is not enabled at the 90478
receiver
■
Be sure transmitter DIP switch 8 is
OFF
SpO
2
value displays ???
■
Sensor not connected to patient
■
Re-attach sensor
■
Adapter cable not connected to module
properly
■
Correctly connect the adapter cable
■
Sensor not connected to adapter cable
■
Correctly connect the sensor
■
Excessive patient motion
■
Urge patient to remain still while
reading is in progress
■
Transmitter is in the initialization phase
(the first 15 seconds after sensor
application)
■
Wait until initialization is complete
■
Low battery indicator constantly
illuminated
■
Call qualified service person
Insufficient signal or
noisy signal
■
Sensor placement not optimum
■
Move sensor to a site which has better
perfusion
■
Align LED with sensor photodetector
■
Sensor placed below blood pressure
cuff
■
Move sensor to an alternate limb
Intermittent or complete
failure to operate
■
Depleted battery
■
Replace battery
■
Low battery light constantly illuminated
■
Call qualified service person
Factors that cause
significant variances in
sensor accuracy
■
Presence of dysfunctional hemoglobins
(COHb, MetHb)
■
Follow hospital procedure for
determining oxygenation in these
patients
■
Presence of intravascular dyes
(indocyamine green, methylene blue)
depending on their concentration in the
blood stream
■
Follow hospital procedure for
determining oxygenation in these
patients
■
High ambient light level
■
Reduce light levels near patient; wrap
sensor with light blocking material
■
Electrosurgical interference
■
Ultraview digital telemetry is contra-
indicated for electrosurgical use
■
Patient is significantly anemic (Hb less
than 5 gm/dl) or patient has received
large amounts of IV solutions
■
Follow hospital procedure for
determining oxygenation in these
patients