071-0774-00 Rev. A
3-11
SpO
2
Troubleshooting Guide
(continued)
Clinical Situation
Possible Cause
Solution
No SpO
2
alarms are
displayed
■
ECG “Leads Off” condition exists
■
Higher priority alarm condition is
present
■
Re-attach ECG lead wires to patient and
resume ECG monitoring to clear
pending ECG alarms
■
Clear current alarm condition and/or
reprioritize SpO
2
alarms of interest in the
Module Configuration Manager
■
When SpO
2
alarms are set ON, all SpO
2
alarm conditions will cause the
parameter value (or ???) to blink
according to the alarm priority set by
using the Module Configuration Manager