Appendix A: Troubleshooting
VirTis BenchTop 2K, 4K and 6K
40
Revision 004, 12/09
© SP Industries, Inc. 2009
If none of these suggestions eliminate the vacuum problem:
1.
Change and flush the oil in vacuum pump.
2.
Disassemble all vacuum tubing connections, clean with a mild alcohol (such
as methanol), apply a light coating of vacuum grease and reassemble.
3.
Check all threaded and welded connections and seal with vacuum sealant.
If poor vacuum persists, or if testing vacuum with a calibrated gauge indicates a good
vacuum reading, the problem is likely related to the vacuum probe. Replace the
vacuum probe tube and retest.
Note:
If this does not resolve your vacuum problem, contact SP Industries for assistance.
Product Melting
The most common reason for product melting is poor vacuum. Adequate vacuum is
required to keep products in a frozen state. Check for a vacuum leak or possible
restrictions in the vacuum hose as described in the previous section.
If the unit is functioning properly but the product is still melting, the product’s freezing
point (eutectic temperature) may be too low to remain in a frozen state. Retest using a
water sample to determine if a mechanical or product-related problem exists.
Summary of Contents for VirTis BenchTop 2K
Page 1: ...SP INDUSTRIES INC Freeze Dryer Operator s Manual VirTis BenchTop 2K 4K and 6K Freeze Dryers ...
Page 2: ......
Page 6: ...iv Revision 004 12 09 SP Industries Inc 2009 ...
Page 10: ......
Page 30: ...Sentry 2 0 Controls VirTis BenchTop 2K 4K and 6K 20 Revision 004 12 09 SP Industries Inc 2009 ...
Page 36: ...Basic Operation VirTis BenchTop 2K 4K and 6K 26 Revision 004 12 09 SP Industries Inc 2009 ...
Page 44: ...Optional Components VirTis BenchTop 2K 4K and 6K 34 Revision 004 12 09 SP Industries Inc 2009 ...