Sentry 2.0 Controls
VirTis BenchTop 2K, 4K and 6K
12
Revision 004, 12/09
© SP Industries, Inc. 2009
2.
Vacuum
The LEDs on the top wave represent the following vacuum setpoints:
LED
Vacuum (mTorr)
Color
1 2000
Red
2 1000
Red
3 500
Amber
4 200
Amber
5 100
Green
6 50
Green
7
≤
20
Green
Button Functions
Up/Down (+/-) Buttons
The
Up/Down (+/-)
buttons allow you to navigate the menus and increase or decrease
setpoint values displayed on the LCD display.
MENU Button
The
MENU
button performs a number of functions. During an alarm condition,
pressing the
MENU
button will temporarily silence (but not clear) the alarm. When
viewing the Synoptic screen, the
MENU
button will display the Main Menu.
The
MENU
button also acts as an enter key, activating the selected option. To return
to a previous menu, highlight the carat (
∧
) at the top left corner of the LCD screen and
press the
MENU
button.
COND Button
The
COND
button activates the refrigeration system. This provides system cooling for
vapor condensing.
VAC Button
The
VAC
button enables the vacuum pump and opens the vacuum brake solenoid
(VBS) valve, allowing the system to be evacuated. Vacuum level control options can
be programmed from the User Options screen.
Note:
If your vacuum pump is not connected to the IEC plug on the rear of the BenchTop unit,
the
VAC
button will not control your pump. See your pump manual for operation instructions.
AUTO Button
The
AUTO
button will start the refrigeration and vacuum systems in a set sequence.
The condenser will first cool to the condenser setpoint. Once the setpoint is achieved,
the vacuum pump will energize and begin evacuating the system.
DEFR Button
The
DEFR
button activates the condenser defrost function. Hot gas defrost is directed
toward the condenser walls to quickly loosen ice.
Note:
Defrost mode is disabled while any other function is on.
Summary of Contents for VirTis BenchTop 2K
Page 1: ...SP INDUSTRIES INC Freeze Dryer Operator s Manual VirTis BenchTop 2K 4K and 6K Freeze Dryers ...
Page 2: ......
Page 6: ...iv Revision 004 12 09 SP Industries Inc 2009 ...
Page 10: ......
Page 30: ...Sentry 2 0 Controls VirTis BenchTop 2K 4K and 6K 20 Revision 004 12 09 SP Industries Inc 2009 ...
Page 36: ...Basic Operation VirTis BenchTop 2K 4K and 6K 26 Revision 004 12 09 SP Industries Inc 2009 ...
Page 44: ...Optional Components VirTis BenchTop 2K 4K and 6K 34 Revision 004 12 09 SP Industries Inc 2009 ...