10.
Plug the vacuum pump into the receptacle marked VACUUM PUMP on the
back of the unit. The receptacle is an IEC universal outlet. This allows you to
control the pump using the Sentry 2.0. The voltage, phase and frequency of
the vacuum pump must match the voltage, phase and frequency specified on
the rear panel of the freeze dryer.
Note:
You can power your vacuum pump from a wall socket if it does not have an IEC
connector, but you will not be able to control the pump from the Sentry 2.0 Controller.
Only a qualified electrician should perform installation of an IEC connector. The pump
used should not exceed the amperage listed on the BenchTop serial tag.
11.
Enable power to the vacuum pump by switching the pump’s circuit breaker to
the on position.
Oil Mist Eliminators
To reduce fumes from the vacuum pump and/or vent the vacuum exhaust externally,
SP Industries recommends the installation an Oil Mist Eliminator (OME).
Related Parts
•
Oil Mist Eliminator, Plastic for Oerlikon Leybold TRIVAC E D2.5E Vacuum
Pump (P/N 414203).
Summary of Contents for VirTis BenchTop 2K
Page 1: ...SP INDUSTRIES INC Freeze Dryer Operator s Manual VirTis BenchTop 2K 4K and 6K Freeze Dryers ...
Page 2: ......
Page 6: ...iv Revision 004 12 09 SP Industries Inc 2009 ...
Page 10: ......
Page 30: ...Sentry 2 0 Controls VirTis BenchTop 2K 4K and 6K 20 Revision 004 12 09 SP Industries Inc 2009 ...
Page 36: ...Basic Operation VirTis BenchTop 2K 4K and 6K 26 Revision 004 12 09 SP Industries Inc 2009 ...
Page 44: ...Optional Components VirTis BenchTop 2K 4K and 6K 34 Revision 004 12 09 SP Industries Inc 2009 ...