![SP Scientific LYOSTAR II Operator'S Manual Download Page 53](http://html1.mh-extra.com/html/sp-scientific/lyostar-ii/lyostar-ii_operators-manual_1338679053.webp)
Rev 012, 04/11
45
© SP Scientific 2011
Chapter
8
Optional Shelf-Latching Kit
Shelf latching kits are available for each system. These kits can be used with the
LyoStar II shelf assembly to create moveable shelf stacks. The primary use of the
shelf-latching system is to facilitate the stoppering of product vials; however, the shelf
latching kits may also be used to increase the distance between shelves when
processing product in large vials.
For example, if you plan to process vials using a three-shelf lyophilizer and find that
you require additional spacing between the shelves; you may use a shelf latching kit
to lock one or two of the shelves in the upright position. Locking a single shelf would
allow for two equally spaced shelves with a greater distance between shelves
(requires an LSLP-3.1, installation of which is detailed below), while locking two
shelves would allow you to process and stopper vials on a single shelf.
Installation
To install the LSLP 3.1 Shelf Latching kit:
Close the chamber door and move all of the shelves into the full up position
1.
with the stoppering switch.
Locate the threaded rods and wing nuts on the top of the shelf assembly and
2.
loosen the nuts.
Slide a plate between the shelf fluid hoses and the shelf assembly so that the
3.
lip of the plate rests on top of the shelf assembly, with the cutout notch around
the tie rod. Repeat for the other side.
Wing Nuts
Summary of Contents for LYOSTAR II
Page 1: ...LYOPHILIZER OPERATOR S MANUAL FTS SYSTEMS LYOSTAR II STANDARD...
Page 2: ......
Page 6: ...iv Rev 012 04 11 SP Scientific 2011...
Page 29: ...FTS Systems LyoStar II Operating Your Lyophilizer Rev 012 04 11 21 SP Scientific 2011...
Page 38: ...Data Management FTS Systems LyoStar II 30 Rev 012 04 11 SP Scientific 2011...
Page 52: ...Optional Isolation Valve FTS Systems LyoStar II 44 Rev 012 04 11 SP Scientific 2011...
Page 58: ...Optional Cleanroom Configuration FTS Systems LyoStar II 50 Rev 012 04 11 SP Scientific 2011...
Page 66: ...General Maintenance FTS Systems LyoStar II 58 Rev 012 04 11 SP Scientific 2011...
Page 70: ...Appendix A Troubleshooting FTS Systems LyoStar II 62 Rev 012 04 11 SP Scientific 2011...
Page 71: ......