Sample Pattern 2 Description
The pattern starts at 22 °C and soaks at that temperature for 10 minutes. The
program then ramps from 22 °C to -5 °C in 11 minutes and then soaks at -5 ° for 30
minutes. A ramp from -5 °C to -30 °C over the next 84 minute period results in a ramp
rate of 0.3 °C/m. The pattern ramps at 0.1 °C/m down to -35 over a period of 50
minutes. The program ends when it reaches -35 °C in step 5.
Note:
The system has been factory programmed to sound an alarm when the programmed
pattern ends.
The trend line of Sample Pattern 2 is shown in green below: