13
Capacitance Test
Operation cont.
1. Set the rotary function switch to
the
Need
position.
2. Insert the black test lead into the
COM
input jack and the red test
lead into the
Need
input jack.
3. Touch the test lead probes to the
capacitor being tested.
4. Read the capacitance value on the LCD display. Large
capacitors may take up to a minute to get a stable reading.
Frequency and Duty Cycle Measurements
1. Set the rotary function switch to the
Hz
position.
2. Press the
Hz %
button to select Frequency (Hz) or Duty Cycle (%).
The
“Hz”
or
“%”
symbol will appear on the LCD display.
3. Insert the black test lead into the
COM
input jack and the red test lead
into the
Hz %
input jack.
4. Touch the test lead probes to the
circuit being tested.
5. Read the frequency or duty cycle on
the LCD display.
6. When in the AC voltage or AC current
function, press the
Hz %
button once,
to select Frequency (Hz) or twice to select Duty Cycle (%).
Press the
Hz%
button again to return to voltage or current
readings. The measurements will be made through the test
leads when the meter is set to voltage and through the clamp
jaws when the meter is set to current. Refer to specifications
in the manual for frequency range and sensitivity.
Safely discharge capacitors
before taking capacitance measurements.
WARNING:
Observe all safety precautions
when working on live voltages.
WARNING:
600A
1000A
MODE
REL
AC/DC TRMS CLAMP METER
AUTO
POWER
OFF
MAX
MIN
23070T
Non-Contact
Voltage Detector
CAT IV 600V
CAT III 1000V
600A
1000A
MODE
REL
AC/DC TRMS CLAMP METER
AUTO
POWER
OFF
MAX
MIN
23070T
Non-Contact
Voltage Detector
CAT IV 600V
CAT III 1000V
14
Temperature Measurements
Operation cont.
1. Set the rotary function switch to the
°F °C
position.
2. Press the
MODE
button to select readings in °F or °C.
3. Connect the Temperature Probe to the Banana Plug Adapter.
Note the – and + markings on the adapter. Connect the adapter
to the meter, making sure the – side goes into the
COM
input
jack and the + side goes into the
°F °C
input jack.
4. Touch the tip of the Temperature Probe to the object being
measured. Keep the probe touching the object until the
reading stabilizes (about 30 sec).
5. Read the temperature on the LCD display.
1. The non-contact voltage detector operates when
the meter is set to any measuring function.
The detector does not operate when Auto Power Off
turns the meter off or when the rotary function switch
is set to the off position.
2. Slowly move the detector probe closer to the conductor
being tested.
3. If AC voltage within the specified range is present, the indicator light will illuminate.
To avoid electric shock, remove the temperature
probe before changing to another measurement function.
WARNING:
Non-Contact AC Voltage Detector
(100 to 1000V AC)
WARNING:
Risk of Electrocution. Before use, always
test the Voltage Detector on a known live circuit to verify proper operation.
WARNING:
Insulation type and thickness, and distance from the voltage source
and other factors may effect operation. Use other methods to verify live voltage if there is
any uncertainty.
NOTES :
The detector is designed with high sensitivity. Static electricity and other sources of
electrical energy may randomly activate the detector. This is normal operation.
The detector only activates the indicator light when AC voltage is present. It does not indicate
the voltage level on the LCD display.
WARNING:
Do not touch the temperature probe to live circuits.