Input Limits
Function
Maximum Input
Amperage AC
200A AC
Voltage AC, Voltage DC
600V AC/DC
Resistance, Diode Test,
Continuity
600V AC/DC
General Specifications
4
Clamp jaw opening
Insulation
Display
Polarity
Over Range Indication
Diode Test
Continuity Test
Low Battery Indication
Measurement Rate
Auto Power Off
Input Impedance
AC Response
AC Bandwidth
Battery
Operating Temperature
Storage Temperature
Relative Humidity
Operating Altitude (Maximum)
Weight
Dimensions
Safety
1” (25mm)
Class 2, Double insulation.
2000 count LCD
Minus symbol “-“ is displayed for negative polarity
“OL” is displayed
Open circuit voltage approx. 2V, Test current approx. 1mA
Beeper on
≤
50
Ω
, Beeper off
≥
70
Ω
“ “ is displayed
3 times per second, nominal
After approx. 15 minutes of inactivity
AC voltage/DC voltage:
≥
10M
Ω
Average responding
AC voltage: 45 to 400Hz, AC current: 50 to 60Hz
Two AAA 1.5V batteries
32° to 104°F (0° to 40°C)
14° to 122°F (-10° to 50°C)
Maximum, non-condensing: 95% up to 82°F (28°C),
75% to 104°F (40°C), 45% to 122°F (50°C)
7000 feet(2100meters)
252g (including AAA batteries)
7.8” x 2.7” x 1.4” (198 x 69 x 36mm)
UL
:
61010 - 1:2012
.
61010-2-032: 2014
.
61010-2-033:
2014 EMC
:
EN61326-1:2013
,
EN61326-2-2:2013
3
International Safety Symbols
Potential danger. Indicates the user must refer to the
manual for important safety information.
Indicates hazardous voltages may be present
Equipment is protected by double or reinforced insulation
Indicates the terminal(s) so marked must not be connected to a circuit where
the voltage with respect to earth ground exceeds the maximum safety rating
of the meter .
Indicates the terminal(s) so marked may be
subjected to hazardous voltages.
MAX
600V
The measurement category (CAT) rating and voltage rating is determined by a combination of the meter, test probes
and any accessories connected to the meter and test probes. The combination rating is the LOWEST of any
individual component.
Brief Description
Typical Applications
Category
Rating
Single phase receptacles
and connected loads
Three phase circuits and
single phase lighting
circuits in commercial
buildings
- Household appliances, power tools
- Outlets more than 30ft (10m) from a CAT III source
- Outlets more than 60ft (20m) from a CAT IV source
- Equipment in fixed installations such as 3-phase
motors, switchgear and distribution panels
- Lighting circuits in commercial buildings
- Feeder lines in industrial plants
- Any device or branch circuit that is close to a CAT III source
CAT II
CAT III
Connection point to utility
power and outdoor
conductors
- Primary distribution panels
- Overhead or underground lines to detached buildings
- Incoming service entrance from utility
- Outdoor pumps
Cat IV
Safety Category Ratings
WARNING:
Operation is limited to CAT II applications when the insulated tips are
removed
from one or both test probes. Refer to Input Limits section of this manual for maximum voltage ratings.