Part 2 Installation

2.7   

Installing the Oxygen Sensor, using H1 Flow KF-40 Sensor Housing

The OMD-501D can accept either a TO2-1x or TO2-2x (CO2 Applications) trace oxygen sensor.  For proper 

selection with your application, please view our online application guide or contact your local sales rep.

Prior to installing the sensor, it is important to make sure that the analyzer gas lines are hooked up and the 

unit is ready to purge with a zero gas.  Connect the zero gas line and set your flow between 0.5 - 4.0 SCFH.  

The sensor housing can also be installed in a low O2 environment with no flow such as a glove box or other 

closed environment.

  

To Install the Sensor:

 

 

 

-  Remove the top portion of the sensor housing by unscrewing it.

 

 

 

-  Lift up on the top of the sensor housing and set to the side.

 

 

 

-  Inspect O’ring for cracking, replace if necessary.  

Always lube your Orings!

 

 

 

-  Remove the sensor from its box.  With scissors, open nitrogen purged packaging  

 

 

 

    and remove the sensor.

 

 

 

-  Visually inspect sensor for damage, if damaged notify the factory immediately.

 

 

 

-  Remove the shorting pins across the back of the sensor board contacts.

 

 

 

-  Place the sensor inside the bottom housing with the metal screen mesh facing down  

 

 

 

    and the circuit board contacts facing up.

 

 

 

-  Return upper portion of the sensor housing to the stainless steel bottom.  Hand    

 

 

 

    tight should be acceptable to create an airtight seal.

 

 

 

-  Immediately insert into glove box or other atmosphere with zero gas.

 

 

 

-  If the analyzer has not been calibrating, refer to section 4.1 for more information.

  

   CAUTION:  Prior to installing the oxygen sensor.  Read section 4.1 

on performing a span calibration.

H1 Sensor Housing

Top portion of sensor housing

Bottom portion of sensor housing

Summary of Contents for OMD-501D

Page 1: ...OMD 501D Trace Oxygen Analyzer 4045 E Guasti Rd 203 Ontario CA 91761 USA Web www sso2 com Phone 1 949 398 2879 E mail sales sso2 com OMD 501D Manual Rev 1 3 1_2015 Instruction Manual...

Page 2: ...nnections 2 6 Installing the Oxygen Sensor using the H3 Flow Through Sensor Housing 2 7 Installing the Oxygen Sensor using the H1 KF40 Sensor Housing 3 1 Understanding the Controls and Their Operation...

Page 3: ...Southland Sensing Ltd under a strict quality proceedure To understand how the oxygen analyzer functions it is important to understand a little bit of the sensor characteristics The active components...

Page 4: ...ight Dimensions 1 4 DIN 96 x 96 x 65mm Enclosure Flush Mount ABS Classification General Purpose Temperature 0 50 deg C Alarms 2 Adjustable NO or NC Power 12 28 VDC or 100 240VAC Signal Output 4 20mA R...

Page 5: ...he factory If damage to any portion of the new analyzer is present stop and report damage to the shipping company as well as the factory The analyzer is shipped with all materials needed to install an...

Page 6: ...ts The OMD 501D is designed as an indoor unit For field installations please refer to the NEMA4X IP66 en closure option offered by Southland Sensing Ltd Refer to part 3 operation section of this instr...

Page 7: ...sensor housing it is important to make sure the bottom portion of the sensor housing is mounted flat and the top portion of the sensor housing which is manufactured of nylon is at the top This will i...

Page 8: ...e polarity of the power be connected in reverse it is recommended to check polarity before running power to the unit It is also recommended to add an external fuse to the power inlet the analyzer does...

Page 9: ...rom its box With scissors open nitrogen purged packaging and remove the sensor Visually inspect sensor for damage if damaged notify the factory immediately Remove the shorting pins across the back of...

Page 10: ...he sensor housing and set to the side Inspect O ring for cracking replace if necessary Always lube your Orings Remove the sensor from its box With scissors open nitrogen purged packaging and remove th...

Page 11: ...tion or SPAN CAL To adjust set or use the delay mode of the two onboard user programmable Alarm Relay Contacts All features are programmable selectable through the MENU bottom The UP DOWN arrows will...

Page 12: ...0 1000 ppm 3 VDC Range 4 0 1 4 VDC Range 5 0 25 5 VDC From the HOME screen press the MENU key and the display will indicate Use the UP DOWN keys to move the cursor to allow the user select AUTO RANGE...

Page 13: ...ercent value based on your alarm set point Press ENT once selected From here you can use the UP DOWN arrows to adjust each digit which is highlighted in red Simply hit the enter button to move the cur...

Page 14: ...can verify the signal matches with the following formula Signal Output mA Reading Full Scale Range x 16 4 For example if we are reading 500ppm on the 1000 ppm range Signal Output mA 500 1000 x 16 4 Si...

Page 15: ...ts are firmly touching the gold pogo pins on the housing The H1 sensor housing alread has a hole at the bottom of the sensor that exposes the sensing surface to ambient air Let the reading stabilize f...

Page 16: ...tood Consult the factory if any questions arise Note For a new sensor purging with a zero gas for 4 6 hours will help the low end stability and response Connect the gas sample line and set the pressur...

Page 17: ...he sensor No tools are required to replace the sensor Simply unscrew Counter Clockwise the collar H3 flow through sensor housing or top portion H1 KF 40 sensor housing Once free open the top portion o...

Page 18: ...ide or Acetic Acid electrolyte Lead anode Handling Rubber or latex gloves safety glasses Storage Indefinitely Physical Properties Boiling Point Range KOH 100 to 115 C or Acetic Acid 100 to 117 C Melti...

Page 19: ...PEL 10 ppm TWA Lead OSHA PEL 05 mg cubic meter Ingestion Electrolyte could be harmful or fatal if swallowed KOH Oral LD50 RAT 2433 mg kg or Acetic Acid Oral LD50 RAT 6620 mg kg Eye Electrolyte is cor...

Page 20: ...pm 0 100ppm 0 1000ppm 0 1 0 25 Power 12 28 V DC 100 240 V AC Alarms 2 Adjustable High Low with Delay Mode Analog Output 4 20mA Isolated Range ID 1 5VDC in 1VDC increments Display Color with Backlight...

Reviews: