21. NON PROGRAMMABLE PARAMETERS
21. NON PROGRAMMABLE PARAMETERS
Interval
Values
Committed period
95 ms (± 5 ms)
Atrial refractory periods
Values
Post atrial sensing
47 ms (± 16 ms)
Post atrial pacing
109 ms (± 4 ms)
Ventricular refractory periods
Values
Post ventricular sensing
95 ms (± 16 ms)
Post ventricular pacing
220 ms (± 4 ms)
Post atrial pacing (blanking)
16 ms (± 3 ms)
Tachycardia criteria
Values
Window of PR association
63 ms (± 1 ms)
Therapies
Values
Waveform
(1)
Constant tilt (50% - 50%)
Stored energy for the Max. shock
42 J
Pacing amplitude during ATP
therapies
7 V (Actual value at 300 ms: 5.3 V )
Anti-PMT protection
Termin
(1) The device has 50% tilt in each phase thus delivers 94% of stored energy. Each phase is limited to
10 ms duration.
SORIN – INTENSIA DR 154 – U152A
59