20. PROGRAMMABLE PARAMETERS
20. PROGRAMMABLE PARAMETERS
measured at 37 °C under a 500 ohm load
Legend:
Value in bold
: "as shipped" value
Underlined value: nominal value
20.1.
ANTIBRADYCARDIA PACING
Basic parameters
Values
Mode
VVI-VVIR-DDD-DDDR-DDD/DDIR-DDI-DDIR-
SafeR (AAI
<=> DDD)
-SafeR-R (AAIR <=> DDDR)
Basic rate (bpm)
(1)
From 30 to 90 by steps of 5 ;
60
(± 4 %)
Maximum rate (bpm)
From 100 to 145 by steps of 5 ;
120
(± 6 %)
Rate hysteresis (%)
0
-5-10-20-35 (± 18 ms)
Rest AV delay (ms)
30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-
155
-165-170-180-190-195-205-210-220-225-235-250
(± 19 ms)
Exercise AV delay (ms)
30-40-45-55-65-70-
80
-85-95-100-110-115-125-135-140-150-
155-165-170-180-190-195-205-210-220-225-235-250
(± 19 ms)
AVD Paced/Sensed Offset (ms)
0-10-15-25-30-40-45-55-
65
-70-80-85-95-100-110-115-125
(± 1 ms)
(1) The corresponding periods are (in ms): 2000-1714-1500-1333-1200-1091-1000-923-857-800-750-
706-667 ms.
Special features
Values
Smoothing
OFF
-Very slow-Slow-Medium-Fast
Mode Switch
ON
-OFF
Physical activity
Very low-Low-
Medium
-High-Very high