15. SPECIAL MODES
15.4.
DETECTION CHARACTERISTICS IN THE PRESENCE OF
ELECTROMAGNETIC FIELDS
Per Clause 27.4 of Standard EN 45502-2-2, detection in the presence of electromagnetic
fields is characterized as follows:
─
Differential mode:
─
Common mode rejection ratio:
16.6 Hz
50 Hz
60 Hz
Atrial channel
≥ 75 dB
64 dB
62 dB
Ventricular channel
≥ 69 dB
≥ 57 dB
≥ 56 dB
Modulated interference:
For atrial sensitivity setting of 0.2 mV, compliance to the Cenelec standard 45502-2-2 is met
for a maximum test signal amplitude of 8 V for the frequency of 60 MHz. 0.4 mV complies
with the standard for the whole frequency range.
15.5.
PROTECTION AGAINST SHORT-CIRCUITS
The defibrillator can undergo a short-circuit if the anode and cathode are not adequately
separated.
In this case, the shock is aborted to prevent damaging the defibrillator and a warning will
indicate that a short circuit (shock impedance < 20 ohms) was detected during the last
shock.
SORIN – INTENSIA DR 154 – U152A
39