background image

 14.  IMPLANT PROCEDURE

14.4.

BEFORE OPENING THE PACKAGE

Before opening the package, check the "Use Before" date printed on the labels on the box 
and on the sterile package. Defibrillators that have not been implanted before that date 
should be returned to Sorin.

Interrogate the device: 

if a warning is displayed, do not implant the device and contact your Sorin 
representative.

if magnet rate is lower than 91 min-1, and if the last reforming/charge occurred more 
than one week ago, do not implant the device. Otherwise wait for one more week before 
checking the magnet rate again. 

NOTE : The magnet rate and battery voltage can decrease before the expiration date is 
reached. However, the magnet rate should be equal to or higher than 91 min-1 at the time of 
implant.

Devices MUST NOT be interrogated and programmed within the vicinity of other devices.

Also check the integrity of the sterile package. The sterility of the contents is no longer 
guaranteed if the package has been pierced or altered. If the defibrillator is no longer sterile, 
it  should   be  returned  in  its  packaging  to  Sorin.  Any re-sterilization   of   the  unit  is  at  the 
discretion of Sorin.

14.5.

PRIOR TO IMPLANTATION

Use the programmer to verify the defibrillator can be interrogated before implantation.

Verify   all   shock   therapies   are   disabled   in   order   to   avoid   accidental   discharge   during 
implantation.

It   is   not   advisable   to   program   the   Smoothing   function   before   implantation,   since   the 
defibrillator may detect noise and pace at a rate higher than the programmed basic rate.

14.6.

DEVICE PLACEMENT

The   pocket   should   be   prepared   in   the   left   pectoral   position,   either   subcutaneously   or 
submuscularly.   Subcutaneous   device   implantation   is   recommended   for   optimal   RF 
communication efficacy.

Implantation in an abdominal position is not advisable.

In its final position, the defibrillator should be no more than 4 cm below the skin surface.

14.7.

CHOOSING THE TYPE OF LEAD

The defibrillator should be connected to:

one bipolar atrial sensing/pacing lead 

one quadripolar/tripolar ventricular defibrillation lead with sensing/pacing bipolar 
electrodes, and one or two defibrillation electrodes.

The choice of leads and their configuration is left to the implanting physician’s judgment.

34

SORIN – INTENSIA DR 154 – U152A

CAUTION:

 Do not shake or tap sharply on the ICD package with the ICD inside, because 

the ICD's sensing circuits can interpret this as P-waves or R-waves and record these as an 
arrhythmia   episode.   If   unusual   shaking   or   tapping   of   the   package   results   in   a   stored 
arrhythmia episode, erase the recording before using the ICD.

Summary of Contents for Intensia DR 154

Page 1: ...Implant manual Implantable cardioverter defibrillator DR model 154...

Page 2: ...blank...

Page 3: ......

Page 4: ...blank...

Page 5: ...atient counselling information 24 9 Conformance to standards 25 10 Physician guidelines 28 10 1 Physician training 28 10 2 Directions for use 28 10 3 Maintaining device quality 28 11 Patient informati...

Page 6: ...olter Function 44 17 3 Recommended Replacement Time RRT 45 17 4 Explantation 45 17 5 Defibrillator identification 46 18 Physical characteristics 47 18 1 Materials used 47 19 Electrical characteristics...

Page 7: ...NSIA DR 154 is also equipped with the RF wireless technology which enables to remotely monitor the patients who have the Sorin SMARTVIEW Monitor installed at home INTENSIA DR 154 provides high energy...

Page 8: ...f the following situations Survival of at least one episode of cardiac arrest manifested by the loss of consciousness due to ventricular tachyarrhythmia Recurrent poorly tolerated sustained ventricula...

Page 9: ...ion digitalis intoxication drowning electrocution electrolyte imbalance hypoxia sepsis or unstable ischemic episodes who present incessant tachyarrhythmia who have an internal pacemaker whose primary...

Page 10: ...Shock Therapy to OFF during surgical implant and explant or post mortem procedures The device can deliver a serious high energy shock should accidental contact be made with the defibrillation electro...

Page 11: ...he cardiac defibrillator Set it at minimum intensity Use it briefly 4 After the procedure check for proper implant function The device should never be exposed directly to the diathermy source External...

Page 12: ...the device or accessories if the packaging is wet punctured opened or damaged because the integrity of the sterile packaging may be compromised Return the device to the manufacturer Device Storage Sto...

Page 13: ...to the lead body tie it too tightly or otherwise create excessive strain at the insertion site as this may damage the lead Use the lead stabilizer to secure the lead lateral to the venous entry site...

Page 14: ...nal paddles adjust the external paddle position e g anterior lateral to anterior posterior and be sure that the external paddle is not positioned over the patch Do not fold alter or remove any portion...

Page 15: ...t were treated with surgery or with external defibrillation of a ventricular cardiac event Observations are defined as symptomatic or asymptomatic clinical events with potential adverse device effects...

Page 16: ...est 2 5 3 2 10 5 Ventricular oversensing 3 7 9 3 15 7 A patient can have more than one observation There were 228 7 device months in this study These observations would not have happened with the curr...

Page 17: ...ious non pacemaker related events 8 17 8 8 1 7 a 4 74 device years Non serious events due to the use of SafeR 2 AAI DDD included delay in switching on 2nd degree AV block inappropriate classification...

Page 18: ...y available defibrillator lead outside the U S Methods Investigators selected patients who survived at least one episode of cardiac arrest manifested by loss of consciousness presumably due to a ventr...

Page 19: ...Upper 95 CI Europe 20 0 77 0 17 0 04 0 69 0 84 Small Centers 9 0 79 0 18 0 06 0 67 0 91 Large Centers 11 0 75 0 15 0 05 0 66 0 84 Males 17 0 77 0 16 0 04 0 70 0 85 Females 3 0 73 0 22 0 13 0 47 0 98...

Page 20: ...study 6 2 SAFER AAI DDD STUDY SafeR AAI DDD mode in INTENSIA is similar to that in Symphony The differences in SafeR AAI DDD mode between the two devices are To prevent long RR intervals during VT VF...

Page 21: ...le Mean patient age SD was 74 9 years Primary indications for implant were 1st degree AV block 11 1 2nd degree AV block 6 7 3rd degree AV block 22 2 sinus node dysfunction 62 2 or other 6 7 Effectiven...

Page 22: ...block brady tachy syndrome 6 and brady tachy syndrome sinus node disease 5 Adverse events Clinical study of the SafeR AAI DDD included 45 Symphony 2550 devices implanted in 45 patients No serious adve...

Page 23: ...V block inappropriate classification of a PAC disagreement between markers and recorded EGM atrial pacing above the maximum rate recycling on an R wave in a refractory period and disagreement in the s...

Page 24: ...h patients Antiarrhythmic drug therapy If the patient is being treated with antiarrhythmic or cardiac drugs the patient should be on a maintenance drug dose rather than a loading dose at the time of I...

Page 25: ...ce there has been no quantitative assessment of the presence of leachables in breast milk Pediatric Patients This device has not been studied in patients younger than 18 years of age Geriatric Patient...

Page 26: ...lowing points in counselling the patient about this device Persons administering CPR may experience tingling on the patient s body surface when the patient s ICD system delivers a shock Advise patient...

Page 27: ...tronic and electrical apparatus with the basic restrictions related to human exposure to electromagnetic fields 10 MHz 300 GHz EN 301 489 1 v1 8 1 EN 301 489 27 v1 1 1 Electromagnetic compatibility an...

Page 28: ...ers used to communicate weather data the Meteorological Satellite or the Earth Exploration Satellite Services and must accept interference that may be caused by such stations including interference th...

Page 29: ...est susceptible de compromettre le fonctionnement du dispositif Conform ment la r glementation d Industrie Canada le pr sent metteur radio peut fonctionner avec une antenne d un type et d un gain max...

Page 30: ...e screen on the Sorin dedicated programmer Paper copies of Online Help can be obtained by contacting your Sorin representative 10 3 MAINTAINING DEVICE QUALITY This device is FOR SINGLE USE ONLY Do not...

Page 31: ...n the outer storage package Additional copies can be obtained by contacting your Sorin representative or on the Sorin s web site http www sorin com This information should be given to each patient wit...

Page 32: ...WARRANTY AND REPLACEMENT POLICY Sorin warrants its defibrillators Refer to the section Warranty for additional information Please see the following labelling sections for information concerning the pe...

Page 33: ...utomatic AV delay algorithm as well as a mode switching function are available INTENSIA DR 154 offers tiered therapy Therapies can be programmed independently in each zone in the Slow VT and VT zones...

Page 34: ...with the IS 1 standard and defibrillation ports is compatible with the DF 4 standard Distal lead terminal connections are secured with set screws accessed via self sealing silicone plugs All lead con...

Page 35: ...e packaging contains a defibrillator and one screwdriver The non sterile items contained in the outer storage package are the implant manual the ICD Registration Form and its envelope the patient book...

Page 36: ...to verify the defibrillator can be interrogated before implantation Verify all shock therapies are disabled in order to avoid accidental discharge during implantation It is not advisable to program th...

Page 37: ...LANT Pacing and sensing thresholds should be measured at implant Pacing thresholds Acute thresholds should be lower than 1 V or 2 mA for a 0 35 ms pulse width both in the ventricle and in the atrium S...

Page 38: ...he lead pin all the way into the port check that the pin protrudes beyond the distal insert 6 Tighten check the tightness and ensure the lead pin still protrudes beyond the distal insert and did not m...

Page 39: ...g the hole provided for this purpose in order to avoid potential migration of the device into the pectoral muscle 14 12 TESTS AND PROGRAMMING During the implant testing procedure it is recommended tha...

Page 40: ...ention Mode Switch Anti PMT Smoothing Rate Response When the magnet is removed the sensor rate is forced to the basic rate arrhythmia detection algorithms and sequential therapies are reinitialized th...

Page 41: ...atrial sensitivity setting of 0 2 mV compliance to the Cenelec standard 45502 2 2 is met for a maximum test signal amplitude of 8 V for the frequency of 60 MHz 0 4 mV complies with the standard for t...

Page 42: ...ventricular tachyarrhythmia onset Searching for a long cycle P And R based Arrhythmia Detection PARAD Additional arrhythmia classification criterion to improve identification of atrial fibrillation an...

Page 43: ...he committed period prevents inappropriate ventricular inhibition if crosstalk occurs Protection against noise Allows the distinction between ventricular noise and ventricular fibrillation If the devi...

Page 44: ...16 7 1 SMARTVIEW Monitor The SMARTVIEW monitor is a small device equipped with an RF transmission module to communicate with the implant and a modem to export data through the internet The SMARTVIEW...

Page 45: ...follow ups of their patient configure additional ways of being notified of alerts for instance by SMS fax or e mail consult print and export patient reports 16 7 5 Alert system The following set of a...

Page 46: ...a device reset the magnet rate is equal to 87 bpm it will be updated within the next 24 hours Refer to the online help for a description of displayed warning and the necessity to contact Sorin for an...

Page 47: ...at the charge time increases Under normal conditions and without programmer use the charge times are as follows Shock energy Charge time sec BOS 3 42 J 10 2 RTT 42 J 13 3 1 Recommended Replacement Tim...

Page 48: ...rogramming head interfaced with the Sorin dedicated programmer Position the programming head over the telemetry antenna located in the upper part of the device in order to communicate effectively via...

Page 49: ...8 6 cm3 Active surface area of casing 76 cm2 Connector Atrium IS 1 Ventricle DF 4 18 1 MATERIALS USED Active surface area of casing 99 pure titanium Connectors Polyurethane and silicone elastomer Medi...

Page 50: ...energies maximum voltages and delivered energies at 37 C 50 ohm load for the minimum low mean and maximum programmed energy values is as follows Stored energy J 0 5 10 20 34 42 V1 Volt 75 341 483 631...

Page 51: ...r G sensor ON 8 7 years Pacing in DDD mode 15 in atrium 1 in ventricle 500 ohm 3 5 V 0 35 ms 60 bpm one 42 J shock per quarter sensor OFF 7 8 years Pacing in DDD mode 15 in atrium 15 in ventricle 500...

Page 52: ...18 ms Rest AV delay ms 30 40 45 55 65 70 80 85 95 100 110 115 125 135 140 150 155 165 170 180 190 195 205 210 220 225 235 250 19 ms Exercise AV delay ms 30 40 45 55 65 70 80 85 95 100 110 115 125 135...

Page 53: ...ar signal of 2 13 ms 2 The correlation between the programmed amplitudes the stored amplitudes and the mid pulse delivered amplitudes under a 500 ohm load are given in the following table Programmed a...

Page 54: ...0 5 0 6 0 75 0 85 1 10 V amplitude V 1 1 5 2 2 5 3 3 5 4 4 5 5 6 20 V pulse width ms 0 12 0 25 0 35 0 5 0 6 0 75 0 85 1 10 Refractory periods Values Atrial refractory period post ventricular sensing...

Page 55: ...0 30 50 100 200 VT persistence cycles 4 6 8 12 16 20 30 50 100 200 VF persistence cycles From 4 to 20 by steps of 1 6 1 The Slow VT zone should be programmed ON only if the VT zone is programmed ON De...

Page 56: ...in slow VT zone ATP 1 program Values ATP program OFF Burst Burst Scan Ramp Ramp Scan Number of sequences 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Cycles in first sequence 1 2 3 4 5 6 7 8 9 10 11 12 13 14 1...

Page 57: ...3 1 5 2 2 5 3 3 5 4 5 6 7 8 9 10 12 14 16 18 20 22 24 26 28 30 32 34 42 Shock 2 J OFF 0 5 0 8 1 1 3 1 5 2 2 5 3 3 5 4 5 6 7 8 9 10 12 14 16 18 20 22 24 26 28 30 32 34 42 Number of Max Shock 42 J OFF 1...

Page 58: ...2 2 5 3 3 5 4 5 6 7 8 9 10 12 14 16 18 20 22 24 26 28 30 32 34 42 Shock 2 J OFF 0 5 0 8 1 1 3 1 5 2 2 5 3 3 5 4 5 6 7 8 9 10 12 14 16 18 20 22 24 26 28 30 32 34 42 Number of Max Shock 42 J OFF 1 2 3...

Page 59: ...ock pacing lead impedance or coil impedance measurements out of range battery depletion etc The Remote tab presents an overview of all the alerts managed by the device General parameters Values RF com...

Page 60: ...AF limit on 24h h 0 5 1 3 6 12 24 Fast V Rate during AT AF ON OFF Fast V Rate limit bpm 80 90 100 110 120 Fast V Duration limit h 0 5 1 3 6 12 24 Therapy information Values Shock disabled ON OFF Shoc...

Page 61: ...16 ms Post ventricular pacing 220 ms 4 ms Post atrial pacing blanking 16 ms 3 ms Tachycardia criteria Values Window of PR association 63 ms 1 ms Therapies Values Waveform 1 Constant tilt 50 50 Stored...

Page 62: ...dioverter defibrillator must be implanted prior to the use before date indicated on the packaging 4 The limited guarantee only applies to suspect devices returned to the manufacturer carefully packed...

Page 63: ...period starting from the implantation date Sorin is committed to replacing free of charge the explanted device by a Sorin device with equivalent features or issuing a replacement credit equal to the...

Page 64: ...41 315 5 776 164 5 776 165 5 818 703 5 836 980 5 868 793 5 891 170 5 891 184 5 899 931 5 931 856 5 935 153 5 954 660 5 978 708 6 181 968 6 230 058 6 236 111 6 251 703 6 256 206 6 307 261 6 337 996 6 3...

Page 65: ...ng ethylene oxide Temperature limitation High voltage Consult instruction for use This icon is used to call your attention to a particularly important point This icon alerts you to a hazard that may r...

Page 66: ...CRM USA Inc Via Crescentino s n 14401 West 65th Way 13040 Saluggia VC Italy Arvada CO 80004 USA Tel 39 0161 487095 Tel 877 663 7674 Facsimile 39 0161 487524 www sorin com 2013 09 U152A CAUTION FEDERA...

Reviews: