9
GB
Recommended separation distances between portable and mobile RF
communications equipment and the UPA-WU10T/UPA-WU10R
The UPA-WU10T/UPA-WU10R is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the UPA-WU10T/
UPA-WU10R can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the
UPA-WU10T/UPA-WU10R as recommended below, according to the maximum output power
of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
IEC 60601-1-2 : 2007
IEC 60601-1-2 : 2014
150 kHz to
80 MHz
d
= 1.2
√
P
80 MHz to
800 MHz
d
= 1.2
√
P
800 MHz to
2.5 GHz
d
= 2.3
√
P
150 kHz to
80 MHz
d
= 1.2
√
P
80 MHz to
2.7 GHz
d
= 2.0
√
P
0.01
0.12
0.12
0.23
0.12
0.20
0.1
0.38
0.38
0.73
0.38
0.63
1
1.2
1.2
2.3
1.2
2.0
10
3.8
3.8
7.3
3.8
6.3
100
12
12
23
12
20
For transmitters rated a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Summary of Contents for UPA-WU10
Page 23: ...主な仕様 23JP ...
Page 61: ...Spécifications 21FR ...
Page 81: ...Technische Daten 21DE ...
Page 247: ...Dane techniczne 23PL ...
Page 269: ...Τεχνικά χαρακτηριστικά 23GR ...
Page 289: ...Technické údaje 21CZ ...
Page 349: ...Technické parametre 21SK ...
Page 405: ...Specificaţii 21RO ...