Guidance and manufacturer’s declaration – electromagnetic immunity
The UPA-WU10T/UPA-WU10R is intended for use in the electromagnetic environment
specified below. The customer or the user of the UPA-WU10T/UPA-WU10R should assure that
it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Electrostatic
discharge
(ESD)
±8 kV contact
±8 kV contact
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, a relative humidity of
at least 30% is recommended.
IEC 61000-4-2
±15 kV air
±15 kV air
Electrical fast
transient/burst
±2 kV for power
supply lines
±2 kV for
power supply
lines
IEC 61000-4-4
±1 kV for input/
output lines
±1 kV for input/
output lines
Surge
±1 kV line(s) to
line(s)
±1 kV
differential
mode
IEC 61000-4-5
±2 kV line(s) to
earth
±2 kV common
mode
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
0%
U
T
(100% dip in
U
T
)
for 0.5/1 cycles
a
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25/30 cycles
a
(for 0.5 sec)
0%
U
T
(100% dip in
U
T
)
for 250/300
cycles
a
(for 5 sec)
Not applicable
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
UPA-WU10T/UPA-WU10R requires
continued operation during power mains
interruptions, it is recommended that the
UPA-WU10T/UPA-WU10R be powered
from an uninterruptible power supply or
a battery.
Summary of Contents for UPA-WU10
Page 23: ...主な仕様 23JP ...
Page 61: ...Spécifications 21FR ...
Page 81: ...Technische Daten 21DE ...
Page 247: ...Dane techniczne 23PL ...
Page 269: ...Τεχνικά χαρακτηριστικά 23GR ...
Page 289: ...Technické údaje 21CZ ...
Page 349: ...Technické parametre 21SK ...
Page 405: ...Specificaţii 21RO ...