5
GB
Important EMC notices for use in medical environments
• The UPA-WU10T/UPA-WU10R needs special precautions regarding EMC and needs to
be installed and put into service according to the EMC information provided in the
instructions for use.
• The UPA-WU10T/UPA-WU10R is intended for use in a professional healthcare facility
environment.
• Portable and mobile RF communications equipment, such as cellular phones, can affect the
UPA-WU10T/UPA-WU10R.
Warning
• Portable RF communications equipment should be used no closer than 30 cm (12 inches)
to any part of the UPA-WU10T/UPA-WU10R. Otherwise, degradation of the performance
of this equipment could result.
• If the UPA-WU10T/UPA-WU10R will be used adjacent to or stacked with other
equipment, normal operation of the UPA-WU10T/UPA-WU10R under such
configurations should be verified via observation.
• The use of accessories and cables other than those specified, with the exception of
replacement parts sold by Sony Corporation, may result in increased emissions or
decreased immunity of the UPA-WU10T/UPA-WU10R.
List of cables used for EMC test
Type of cable
Specifications
USB cables (1-757-429-21) (2)
1 m, Shielded cable
Guidance and manufacturer’s declaration – electromagnetic emissions
The UPA-WU10T/UPA-WU10R is intended for use in the electromagnetic environment
specified below. The customer or the user of the UPA-WU10T/UPA-WU10R should assure that
it is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The UPA-WU10T/UPA-WU10R uses RF energy only
for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The UPA-WU10T/UPA-WU10R is suitable for use in
all establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Summary of Contents for UPA-WU10
Page 23: ...主な仕様 23JP ...
Page 61: ...Spécifications 21FR ...
Page 81: ...Technische Daten 21DE ...
Page 247: ...Dane techniczne 23PL ...
Page 269: ...Τεχνικά χαρακτηριστικά 23GR ...
Page 289: ...Technické údaje 21CZ ...
Page 349: ...Technické parametre 21SK ...
Page 405: ...Specificaţii 21RO ...