4.
Remove the four hexagon cap screws and the four spring washers to detach the front panel.
HEXAGON 3 x 8
SW3
Front panel
Mark
Setscrew
Filter select knob
Knob shaft
Filter with the
lightest color
(1: clear)
Note
• Attach the filtr select knob using the following steps.
(1) Turn the knob shaft to a position where the filter with the lightest color (1: clear) appears when viewed
from the lens mount section.
(2) Match the filter select knob number 1 with the mark on the front panel, and secure the knob with a setscrew.
• Apply threadlocker to the setscrew, then tighten them.
5.
Remove the three screws and remove the transportation case from the CCD unit for repair (option).
P3 x 6
Transportation case
CCD unit for repair
6.
Install the removed parts by reversing the steps of removal.
Work after replacement
Perform the following after the CCD unit is replaced.
• Refer to
• Refer to
• Refer to
HSC300RF/HSC100RF/HSC300R/HSC100R
2-12