ISONIC utPod from Sonotron NDT - Operating Manual – Revision 1.18 - Page 71 of 94
7.2.2.2. Automatic Calibration
Zero to 1
st
Place probe onto the sample of material with known thickness allowing receiving of several back echoes,
then provide gating (
Gate A
) according to paragraph 6.2.1 of the operating manual, then click on
,
the screen as below appears:
Key in the Thickness of the material, on which
calibration is performed
Click on upon completed
Click on
to use automatically determined
USVelocity
and
Probe Delay
value as default:
Multiecho
Refer to paragraph 6.2.2.2 of the operating manual