S-50 DC 

● S-80 DC ● S-110 DC ● S-120 DC – USER MANUAL 

17 

Maintenance and repair 

 

The repair work which can be done on the S-50 / 80 / 110 / 120 DC by the user is limited to the 
replacement of fuses and lamps. Fuses, lamps and tools are included in the service kit. 

 

Repairs to the high-voltage unit are not permitted. 

 

No more testing must be done if the oil level drops. 

 

If the HV test set is defective, the complete set must be sent in for repair, as the operating unit is 
matched to the high-voltage unit (interchangeability not guaranteed). 

 

The S-50 / 80 / 110 / 120  DC is test equipment and as such must be handled and looked after 
with care. 

 

Moisture

, external soiling, direct sunlight and ambient temperatures in excess of 55°C must be 

avoided. 

 

 

NOTE! 

 

The high-voltage unit must only be transported in an upright position. 

 

Technical specifications 

a)  power supply (depending on version) ..............................................................230 V 

± 10% (50 Hz)  

 ........................................................................................................................ 115 V 

± 10% (60 Hz) 

b)  power consumption ..................................................................................................... max. 900 VA 
c)  output voltage ............................................................................. negative polarity, infinitely variable 

  S-50 DC ......................................................................................................................... 0...50 kV 

  S-80 DC ......................................................................................................................... 0...80 kV 

  S-110 DC  .................................................................................................................... 0...110 kV 

  S-120 DC ....................................................................................................................  0...120 kV 

d)  output current 

  S-50 DC ............................................................................................................................... 6 mA 

  S-80 DC ............................................................................................................................... 5 mA 

  S-110 DC ............................................................................................................................. 4 mA 

  S-120 DC .......................................................................................................................... 3.5 mA 

e)  overcurrent tripping 

  S-50 DC ............................................................................................................................. 

8 mA 

  S-80 DC ............................................................................................................................. 

7 mA 

  S-110 DC ........................................................................................................................... 

6 mA 

  S-120 DC ........................................................................................................................ 

5.5 mA 

f)  voltage measuring range 

  S-50 DC  ........................................................................................................................ 

0…50 kV 

  S-80 DC  ........................................................................................................................ 

0…80 kV 

  S-110 DC  .................................................................................................................... 

0…110 kV 

  S-120 DC  .................................................................................................................... 

0…130 kV 

g)  current measuring range 

  ...................................................................................................... 0.1 mA / 1 mA / 10 mA 

 ...................................................................................................................... 

0…9.999 mA 

h)  measuring accuracy 

 .................................................................................. 

±2% (temperature range 23ºC ± 2K)  

 .................................................................................. 

±1% (temperature range 23ºC ± 2K) 

 .............................................................................................. 

±5% (temperature range -25…+55ºC) 

Summary of Contents for S-110 DC

Page 1: ......

Page 2: ......

Page 3: ...USER MANUAL HIGH VOLTAGE INSULATION TESTERS S 50 DC S 80 DC S 110 DC S 120 DC SONEL S A Wokulskiego 11 58 100 widnica Version 1 02 11 09 2020...

Page 4: ...S 120 DC USER MANUAL 2 S 50 80 110 120 DC is a modern easy in use and safe testing device Please acquaint yourself with this manual in order to avoid measuring errors and prevent possible problems in...

Page 5: ...g the discharge rod 8 3 4 Attaching the corona ring 9 3 5 Connecting the test equipment 10 4 Operating the device 13 4 1 Controls and displays 13 4 2 Operating procedure 14 4 2 1 Setting the operating...

Page 6: ...th Only operate the system if it is in technically perfect condition No non original parts may be used for the system and its peripherals as the necessary safety will not be guaranteed No mode of work...

Page 7: ...e equation I U R where U measuring voltage I measured current A special protective ground circuit ensures a high level of safety 2 1 Design of the set The test set consists of the operating unit and t...

Page 8: ...80 110 120 DC is used for electrical testing of power cables and plant with direct voltage Areas of application include testing at re commissioning of cable systems after maintenance and repair testin...

Page 9: ...ting unit 3 2 Operating requirements The carrying handle of the operating unit is designed for one hand operating After putting down the device carrying handle can be locked in 30 steps by using a pul...

Page 10: ...ve instructions it must be placed in close proximity to the test object so that the ground and high voltage cables can be kept as short as possible When the equipment is set up in the field it must be...

Page 11: ...spacer is only plugged onto the HV unit without screwing Connect the HV lug only hand tight with the knurled screw Consider following differences to the standard design The smoothing of electricals fi...

Page 12: ...tablish a conductive connection between the station auxiliary ground and the ground terminal of the high voltage unit Screw the high voltage attachment onto the high voltage unit and connect the high...

Page 13: ...gle core system or a three core cable with separate shield on each core All three cores can be tested at the same time Fig 3 4 Connecting the test equipment for insulation testing of a three core cabl...

Page 14: ...ANUAL 12 Fig 3 5 Connecting the test equipment for sheath testing of a 3x single core system The sheaths of all three cores can be tested at the same time The cable shields at the other end of the cab...

Page 15: ...f it is in the leftmost position while enabling the button Unit on lamp white Lights up when the unit is switched on with the mains switch and mains voltage is connected Ground terminal lamp yellow Li...

Page 16: ...ble 1 ground cable for operating unit to station ground Pilot lamp signals proper connection of the station and protective grounds independently of the high voltage being connected The equipment will...

Page 17: ...s exceeded by approx 5 kV the overvoltage trip will re spond If this happens the high voltage will have to be switched back on set the high voltage adjuster to zero first Fig 4 2 Left meter display Ri...

Page 18: ...t must only be operat ed in areas which have been secured and or cordoned off and marked in accordance with EN 50191 and EN 50110 1 standards In an emergency the high voltage must be switched off by a...

Page 19: ...re external soiling direct sunlight and ambient temperatures in excess of 55 C must be avoided NOTE The high voltage unit must only be transported in an upright position 6 Technical specifications a p...

Page 20: ...18 5 kg S 110 DC 20 0 kg S 120 DC 20 5 kg o degree of protection as per EN 60529 operating unit IP40 high voltage unit IP65 high voltage unit at terminals IP00 p max voltage discharge energy for disc...

Page 21: ...e 3 blue lead for high voltage unit length 2 m cable 4 ground cable for discharge rod discharge rod case for discharge rod user manual service kit packing 7 2 Optional accessories Additionally the fol...

Page 22: ...ermal imagers o luxmeters The Calibration Certificate is a document that presents a relation between the calibration standard of known accuracy and meter indications with associated measurement uncert...

Page 23: ......

Page 24: ......

Reviews: