STEP 5
GET CONFIGURATION
OPERATING THE EQUIPMENT
8
If a
Release Mechanism
test is selected the iWAND will display what state the
release is in.
9
Select a position to move the release mechanism to then press
ENT
.
10 The display will show the release mechanism is
Active
.
11 When the release has finished moving, the iWAND will display the new state.
12 Confirm the release mechanism has operated.
13
sequence will ask if a
Sensor
Test is required.
14 If
Yes
is selected the iWAND will display the current measurements of each sensor
available in the instrument.
If the iWAND detects sensors are fitted to the instrument the
Get Configuration
The sensor value will change colour depending on the status of the
measurement. Green:- the measurement is valid and less than 10
seconds old; Orange:- the measurement is greater than 10 seconds old;
Red:- the measurement is reported in error (out of range).
2
NOTE
15 Once all tests have been completed. The iWAND will display
Success Got
Configuration
and the instrument address.