STEP 4
GET CONFIGURATION
The iWAND requires no pre-programming with information. It will automatically discover
the address and download the instrument status.
1
From the iWAND main menu select
Get Configuration
.
2
Hold the iWAND acoustic transducer against the transducer of the instrument.
OPERATING THE EQUIPMENT
The iWAND acoustic transducer must remain in contact with the
transducer of the instrument during any transferral of information. If
contact is not maintained the information transferral will be lost.
2
NOTE
3
Press the
ENT
button on the iWAND to start the
Get Configuration
sequence.
4
The
Discovering
sequence will gather all the information from the instrument.
5
When the iWAND has found the instrument, the address will appear. If this fails,
refer to ‘
Pre-requisites for Operation
’ in
Section 5
of the
iWAND UM-8315
manual.
6
The iWAND will begin to interrogate the instrument for it’s configuration.
7
If the the iWAND detects a
Release Mechanism
is fitted to the instrument the
Get
Configuration
sequence will ask if a Release Test is required.