A Measurement trigger
Measurements are initiated by one of the options listed in the table below.
The commands to trigger measurements via RS-485 and SDI-12 are described in
and
, respectively.
Measured data are either returned directly after the measurement or can be requested by com-
mands via the RS-485 or SDI-12 interface. The format of the returned data can be configured in the
sub-menu
Option
Description
Interval (default)
Measurements are initiated in a specified interval.
TRIG input
Measurements are triggered by the positive edge of a DC-voltage signal
applied to the TRIG input (low: 0 ... 0.6 V, high: 2.2 ... 28 V, min. pulse width:
30 ms)
SDI-12/RS-485
Measurements are externally triggered by commands via RS-485 or SDI-12
from ,e.g. a data logger.
all allowed
Measurement is triggered by all options mentioned above.
B Measurement Interval
An internal measurement interval can be set for the SPA-2. If selected in menu item
, measurements are performed in the defined interval. However, a measurement is always
completed before a new one is initiated.
Unit
sec
seconds
Value range
00:00:01 … 12:00:00
00:01:00 (default)