Thymatron
®
System IV Instructions for Use
57
REPRINT THE LAST TREATMENT
The Thymatron® System IV automatically stores the last treatment in memory. If the recording paper runs out or
is damaged, the last treatment can be reprinted using this feature. This “REPRINT” feature also allows for the
reprinting of the treatment data
just
transferred to the Thymatron® System IV from the PC. In either case, the
only treatment reprinted will be the last one in the memory.
Please note, if the Thymatron printer is re-started after the last treatment is printed and it runs for over thirty
(30) seconds, that data will be considered the last treatment and the previous treatment just printed before the re-
start will be deleted from memory. Also, if the unit is turned off, another treatment is transferred from the
computer or another treatment is given, the earlier treatment data will be lost.
The procedure for reprinting the
treatment data is:
FLEXDIAL™ DATA OUT: REPRINT
While the treatment data is being reprinted, the “REPRINT” message will flash until all the data has been
reprinted, after which the printer will automatically stop, providing the original data contained an end-of-
treatment report.
SET DATE & TIME IN PRINTED REPORT
To change/reset the time or date, use the following procedure. Enter the
FlexDial™
mode,
rotate
to the
“CLOCK” function level and then alternate
pressing
and
rotating
the
FlexDial™
to make the following
selections:
FLEXDIAL™ CLOCK: MONTH DAY YEAR HOUR MIN
MONTH
01 - 12
DAY
01 - 31
YEAR
00 - 99
HOUR
00 - 24
MIN
00 – 60