Levelogger Series User Guide
Page 27
5 Levelogger Series Setup
Depending on the type of datalogger, there will be differences in programming options, memory, and battery
life. See below for a summary of the differences:
Table 5–1: Levelogger Series Programming Options. Memory & Battery Life
Datalogger Type
Sampling Options
Units of Measure
Memory Capacity
Battery Life
Levelogger 5
(Barologger 5)
Linear (0.125 second to 99 hours),
Event Based, Schedule, Repeat
Schedule, Real Time View, Future
Start/Stop
Level: m (default), cm,
ft, kPa, bar, and psi
Baro: kPa (default),
mbar, and psi.
Temp: ºC, ºF
150,000 sets of readings
in Slate or Continuous
Mode
10 years based on
1 reading per minute
Levelogger 5 Junior
Linear (0.5 second to 99 hours),
Real Time View, Future Start
Level: m (default), cm,
ft, kPa, bar, and psi
Temp: ºC, ºF
75,000 sets of readings
in Slate or Continuous
Mode
5 years based on
1 reading per minute
Levelogger 5 LTC
Linear (2 seconds to 99 hours),
Event Based, Schedule, Repeat
Schedule, Real Time View,
Future Start/Stop
Level: m (default), cm,
ft, kPa, bar, and psi
Temp: ºC, ºF
Cond: mS/cm, µS/cm
100,000 sets of readings
in Slate or Continuous
Mode
8 years based on
1 reading every
5 minutes
Rainlogger 5
Event Based (records tips from
tipping-bucket rain gauge),
Real Time View, Future Start
mm or in
Up to 100,000 tip
time logs in Slate or
Continuous Mode
10 years based on
2 parameters logged
every 10 minutes
Levelogger Edge
(Barologger Edge)
Linear (0.125 second to 99 hours),
Event Based, Schedule, Repeat
Schedule, Real Time View, Future
Start/Stop
Level: m (default), cm,
ft, kPa, bar, and psi
Baro: kPa (default),
mbar, and psi.
Temp: ºC, ºF
40,000 sets of readings
in Slate or Continuous
Mode
10 years based on
1 reading per minute
Levelogger Junior Edge
Linear (0.5 second to 99 hours),
Real Time View, Future Start
Level: m (default), cm,
ft, kPa, bar, and psi
Temp: ºC, ºF
40,000 sets of readings
in Slate or Continuous
Mode
5 years based on
1 reading per minute
LTC Levelogger Edge
Linear (2 seconds to 99 hours),
Event Based, Schedule, Repeat
Schedule, Real Time View,
Future Start/Stop
Level: m (default), cm,
ft, kPa, bar, and psi
Temp: ºC, ºF
Cond: mS/cm, µS/cm
27,000 sets of readings
in Slate or Continuous
Mode
8 years based on
1 reading every
5 minutes
Rainlogger Edge
Event Based (records tips from
tipping-bucket rain gauge),
Real Time View, Future Start
mm or in
Up to 60,000 tip
time logs in Slate or
Continuous Mode
10 years based on
2 parameters logged
every 10 minutes
Levelogger Gold
(Barologger Gold)
Linear (0.5 second to 99 hours),
Event Based, Schedule,
Real Time View, Future Start/Stop
Level: m (default),
cm, ft,
Temp: ºC
40,000 sets of readings
in Slate or Continuous
Mode
10 years based on
1 reading per minute
Levelogger Junior
Linear (0.5 second to 99 hours),
Real Time View, Future Start
Level: m (default),
cm, ft,
Temp: ºC
32,000 sets of readings
in Slate Mode
5 years based on
1 reading per minute
LTC Levelogger Junior
Linear (5 second to 99 hours),
Real Time View, Future Start
Level: m (default),
cm, ft,
Temp: ºC
Cond: mS/cm, µS/cm
16,000 sets of readings
in Slate Mode
5 years based on
1 reading every 5
minutes
Rainlogger
Event Based (records tips from
tipping-bucket rain gauge),
Real Time View, Future Start
mm or in
Up to 60,000 tip time
logs in Slate Mode
10 years based on
2 parameters logged
every 10 minutes