S E C T I O N 1 0
•
T r o u b l e s h o o t i n g
36
Level
1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
Slow Flow Rate Troubleshooting
Pressure
Chambers not
activated or
Pressure Cuff
does not inflate
• Pressure Chamber Toggle Switch(es) should be in the
plus (+) pressurized position.
• Pressure Chamber door must be closed with the latch
secured.
• Check to see if the Pressure Regulator Knob is properly
adjusted.
• Check the 1/4” black pneumatic tubing; make sure it is
properly connected and fully seated.
• Check to see if the Fluid Warmer is properly installed
onto the Base Assembly.
• Check to see if the unit is plugged into a working MAINS
receptacle, and power is turned ON.
• Make sure the valve handle on Adapter Assembly is in
the plus (+) pressurized position, not the minus (-) vent
position.
Problem
Check the following
Fluid bag not fully
spiked or bag port
twisted
Make sure the membrane of the fluid bag port is fully
pierced by the bag spike and that the neck of the bag
port is not twisted.
Clamps partly
engaged
Verify all clamps are fully open.
Clamps left in the
clamped position for
long periods
Leaving clamps fully clamped for long periods will cause
the tubing to become deformed. Do not leave clamps
closed for extended periods of time.
Kinked tubing
Verify that no tubing kinks are present.
Problem
Check the following