S E C T I O N 7
•
C o n t r o l s a n d D i s p l a y s
18
Level 1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
Reservoir Level Display
The Reservoir Level Display has a clear window for viewing the amount
of recirculating solution present in the reservoir. Check the reservoir to
ensure the solution level is near the maximum level indicator (
a
). If the
recirculating solution level is too low, the Add Recirculating Solution
LED on the Display Panel illuminates and an audible attention signal
activates.
Pressure Chamber Control Panel
1 Pressurize / Unpressurize Toggle Switch
This switch is used to pressurize the Pressure Chamber.
a
To Pressurize the Pressure Chamber
With Pressure Chamber door closed and latched, push Toggle
Switch all the way to the plus (+) pressurized position. This
applies pressure in the Pressure Chamber when the Fluid Warmer
is turned ON.
b
To Unpressurize the Pressure Chamber
To release pressure from the Pressure Chamber, push Toggle
Switch all the way to the minus (–) unpressurized position.
Pressure is released on the fluid bag in the Pressure Chamber.
2 Pressure Gauge
When the Toggle Switch is in the plus (+) pressurized position
and the Fluid Warmer is ON, this gauge displays the operating
pressure in the Pressure Chamber.
a