![Smiths Medical Level 1 NORMOFLO H-1129 Operator'S Manual Download Page 28](http://html.mh-extra.com/html/smiths-medical/level-1-normoflo-h-1129/level-1-normoflo-h-1129_operators-manual_1305994028.webp)
S E C T I O N 8
•
O p e r a t i o n
Level 1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
23
Alarm Test Mode
The Alarm Test mode is used to test the visual and audible indicators of
the Fluid Warmer. To enter Alarm Test Mode, press and hold the Alarm
Test button (
d
) on the Fluid Warmer's Control Panel.
Mode characteristics
• All visual indicators on the Fluid Warmer's Display Panel (
e
)
illuminate.
• The Fluid Warmer's audible alarm activates.
• When the Alarm Test button is released, the Over Temperature
LED and audible alarm remain active.
To clear this mode:
• Turn the Fluid Warmer OFF and then back ON.
Over Temperature Test Mode
The Over Temperature Test mode is used to test the operation of the
Fluid Warmer's Over Temperature Circuitry. To enter this mode, press
and hold the Over Temperature Test button (
f
) on the Fluid Warmer
Control Panel with the Fluid Warmer at operating temperature (41°C).
Mode characteristics
• The red Over Temperature LED (
g
) illuminates.
• An audible warning signal activates.
To clear this mode
• Turn the Fluid Warmer OFF and then back ON.
Temperature Display
The Temperature Display functions when the Fluid Warmer is powered
ON. Temperature is displayed in degrees Celsius.
d
e
f
g