7
QUICK CHANGE OF DISPOSABLE,
SAME INFUSION VOLUME, SAME PROGRAMMING
1.
With the pump switched on, stop delivery by pressing the
Stop/Start
key
Tasks
02
.
2.
Attach the new disposable. The pump displays the question
Reset
reservoir volume to XX mL?
Press
Yes
to confirm.
3.
Follow the instructions for priming the tubing.
4.
The question
Start the Pump?
appears.
Select
Yes
if you want to restart the previous programming.
1.
Switch off the pump by pressing the
Power on/off
key
.
2.
Remove the disposable.
3.
Attach the new disposable.
4.
Switch on the pump.
5.
Press the button just below
Tasks
.
6.
Using the key, scroll down until the white box highlights
View
Delivery Settings
then press
select
.
7.
Using the key, scroll down until the white box highlights
Reservoir
Volume
then press
select
.
8.
Enter the security code.
9.
Next press or to select the new reservoir volume then press
Save
.
10.
Press the
Stop/Start
key, then select
Review
. Confirm all of the settings
and then press
Next
.
11.
The question
Start the Pump?
appears.
Select
Yes
if you want to restart the previous programming.
CHANGE OF DISPOSABLE WHILE MODIFYING
THE INFUSION VOLUME
Tasks
02