6
ATTACHING THE CASSETTE / ERA
1.
Clamp the tubing, make sure the cassette latch
is unlocked and open the cassette latch to 90
degrees.
2.
Insert the cassette hooks into the hinge pins on
the bottom of the pump.
3.
Without holding the cassette latch, push up on
the cassette until it firmly clicks into place.
4.
Lift the cassette latch into the closed position.
5.
Insert the pump key and turn it clockwise
to lock the cassette if using PCA mode.
6.
When set to PCA mode you must lock the cassette
in order to be able to start delivery. To lock the
cassette, insert the key into the cassette/keypad lock
and turn clockwise.
Correct Incorrect
An uneven gap indicates
that the cassette is not
properly latched.
REMOVING THE CASSETTE OR TUBING
1.
Make sure that delivery has stopped and clamp the tubing.
2.
If the cassette is locked, insert the key and
turn it counter-clockwise until it is in the
unlocked position.
3.
Push down on the cassette latching lever
until the cassette detaches.