SETTINGS TO BE PROGRAMMED:
Continuous Rate:
Specify the drug delivery rate. If the prescription
does not require the continuous rate, enter 0.
PCA Dose:
Amount of drug delivered by pressing the PCA
Dose key or the Remote Dose Cord Button.
PCA Lockout:
Minimum time that must elapse between the time
one PCA dose starts and the time that the next PCA
dose is available.
Delivery Limit:
This option is used to restrict the amount of drug
delivered to the patient (including continuous
+ PCA doses, but not clinician boluses) within
a specified time frame (between 1 hour and 12
hours).
Max Doses/Hr:
Maximum number of PCA doses allowed within a
1-hour period.
KVO Rate:
The KVO (Keep Vein Open) function allows the
delivery of a minimal amount of drug to help
maintain catheter patency.
Reservoir Vol.:
Reservoir volume (cassette or bag).
Start Time:
If a delayed start is programmed, the delivery start
time is indicated here.
PCA / PCEA mode
250 mL
Stopped
Exit
Accept Value
'Accept Value', or press 'select' to edit
Continuous Rate
10
mL/hr
PCA Dose
5
mL
PCA Lockout
30
Min
KVO Rate
0.
1
mL/hr
PCA
11:49 AM
PCA - Epidural
Bupivacaine (0.0625%)
mL
Purple screen
PCA mode
Yellow screen
Epidural PCA