4
I
ntroductIon
Introduction
This Technical Manual is applicable only to the model 2120 CADD
®
-Solis VIP ambulatory infusion pump.
It is intended to provide a basic but limited understanding of the mechanical and electrical operation
of the CADD
®
-Solis VIP ambulatory infusion pump to people familiar with the device. It also outlines
cleaning and functional testing procedures that can be performed on the pump. The CADD
®
-Solis VIP
Operator’s Manual should be used in conjunction with this manual for complete information.
IMPORTANT NOTICE:
CADD
®
-Solis VIP ambulatory infusion pump operations and safety features are based on a
microcomputer design. Inadequate servicing or tampering with the safety features of the pump may
seriously affect performance and safety. For that reason, ALL SERVICING AND REPAIR OF THE
CADD
®
-SOLIS VIP AMBULATORY INFUSION PUMP MUST BE PERFORMED BY SMITHS
MEDICAL OR ITS AUTHORIZED AGENTS.
The manufacturer’s warranty agreement shall become null and void if the pump is not used in accordance
with the Operator’s Manual and Instructions for Use provided with the pump accessories; or if the pump
is serviced by anyone other than Smiths Medical or those authorized by Smiths Medical.
Limited Warranty
The limited warranty associated with the CADD
®
-Solis VIP ambulatory infusion pump can be found in
the product literature supplied with the product when originally purchased, which is incorporated herein
by reference. SMITHS MEDICAL SPECIFICALLY DISCLAIMS ANY OTHER WARRANTY, WHETHER
EXPRESS, IMPLIED OR STATUTORY, INCLUDING, WITHOUT LIMITATION ANY IMPLIED
WARRANTY OF MERCHANTABILITY OR FITNESS FOR USE. Smiths Medical further disclaims
any responsibility for the suitability of the system for a particular medical treatment or for any medical
complications resulting from the use of the system.
The manufacturer shall not be responsible for any incidental damages or consequential damages to
property, loss of profits, or loss of use caused by any defect or malfunction of the system.
If you wish to receive additional information about the extent of the warranty on these products, contact
your Smiths Medical representative or call Customer Service at 1 800.258.5361 (USA) or +1 214.618.0218.
All recommendations, information, and literature supplied by Smiths Medical with respect to the
CADD
®
product line are believed to be accurate and reliable, but do not constitute warranties. No agent,
representative, or employee of Smiths Medical has authority to bind Smiths Medical to any representation
or warranty, expressed or implied.
Warnings
•
The user should ensure that the performance offered by the pump is fit for the intended use and that the
pump is not used in any way or for any purpose other than its intended use.
•
If the pump is dropped or hit, inspect it for damage. Do not use a pump that is damaged or not
functioning properly. Contact Smiths Medical Customer Service to return a pump for service.
•
Do not use rechargeable NiCd or nickel metal hydride (NiMH) batteries. Do not use carbon zinc
(“heavy duty”) batteries. They do not provide sufficient power for the pump to operate properly.
•
Always check the battery compartment for fluid or debris before inserting the batteries, and do not allow
any fluid or debris to fall into the battery compartment. Fluid or debris in the battery compartment may
damage the battery contacts, and could result in loss of power and nondelivery of drug.
•
Residential/facility wiring must comply with all applicable electrical codes. Do not bypass power cord
connections. Do not remove a prong from the power cord.
•
Ensure that the ± 6% system delivery accuracy specification is taken into account when programming
the pump and/or filling the reservoir. Failure to do so may result in medication in the reservoir
becoming depleted sooner than expected.