atterIes
The following tables may be used to predict typical alkaline battery and CADD
®
-Solis rechargeable battery
pack life at different delivery rates. As expected, battery life decreases as the delivery rate increases.
Alkaline battery
life with screen
backlight intensity
set to 3
These estimates are based on laboratory tests conducted at room temperature
using new batteries (Duracell
®
PC1500 / MN 1500, IEC LR6). Actual battery life
varies depending on the battery brand, shelf life, temperature conditions,
delivery rate, and frequency of screen display and backlighting. It is
recommended that new batteries be kept available for replacement.
PCA and Continuous
delivery
(Max. delivery rate
= 100 mL/hr, when
using the PCA
delivery mode)
Delivery Rate
(mL/hr)
Operating Time
(hrs)
Vol. Delivered
(mL)
0.4
142
56
1
.0
139
139
5
.0
124
620
10
.0
113
1130
30
.0
69
2070
50
.0
59
2950
125
.0
37
4625
200
.0
29
5800
350
.0
15
5250
500
.0
11
5500
Intermittent delivery
Volume
(mL)
Duration
(hr)
Cycle
(hr)
KVO
(mL/hr)
Operating Time
(hrs)
Vol. Delivered
(mL)
20.2
1
4
0.2
131
684
40.7
.0
1
4
0.2
116
1221
61
.0
1
6
0.2
111
1177
125
.0
1
6
0.2
75
1637
200
0
1
12
0.2
111
2020
Step delivery
Volume
to Deliver
(mL)
Starting
Rate
(mL/hr)
Step
Duration
(min)
Step Rate
Increase
(mL/hr)
Max Rate
(mL/hr)
No.
of
Steps
Operating
Time
(hrs)
900
45
15
45
315
7
23
1500
37.5
30
80
300
5
24
2500
30
30
90
300
4
19
Taper delivery
Volume
(mL)
Period
(hr)
Taper Up
(hr)
Taper
Down
(hr)
KVO
(mL/hr)
Operating
Time
(hrs)
Vol.
Delivered
(mL)
2000
12
1
1
5
31
5800
3000
12
1
1
5
23
6460