Annual Inspection and Testing Procedures
Inspection Recommendation
Smiths Medical recommends that the following inspection and test procedures be performed annually to
verify function and accuracy of the CADD
®
-Solis VIP pump.
Note:
•
Become familiar with the CADD®-Solis VIP pump before performing the following tests and
procedures. Read the Operator’s Manual supplied with the pump before proceeding.
•
Run the test procedures in manual mode.
•
Many of the tests include steps to unlock the keypad. This is necessary only if the keypad is not already
unlocked.
CAUTION: CADD® pumps are sealed units. A broken or damaged seal will therefore be considered
conclusive evidence that the pump has been misused and/or altered, which voids any and all
warranties. All service and repair of CADD® pumps must be performed by Smiths Medical or its
authorized agents.
I. Visual Inspection
☐
Visually inspect the pump for any damage to the LCD, occlusion sensor seals, valves and
expulsor, cassette hinge area, latch handle, USB connector, lock, 3 cassette sensors, keypad,
indicator lights, AC power jack, remote dose jack, air detector sensor, and housings.
☐
Check the battery door for proper operation. It should not be cracked or broken. Inspect the
battery door seal for proper position.
☐
Examine the battery compartment for damage or debris.
II. Mechanical Inspection
Equipment needed
• 1 CADD
™
medication cassette reservoir
or
1 CADD® administration set
• 1 CADD® key
Procedure
☐
Press each key on the keypad. Each key should have a distinctive tactile feel, and should not feel
flat.
☐
Close and latch the battery door. It should fit snugly in place when closed on the pump.
☐
Attach a CADD
™
medication cassette reservoir or a CADD® administration set to the pump.
Check for smooth operation of the latch. Gently twist and pull on the cassette to make sure it is
firmly attached to the pump.
☐
Lock the cassette by inserting a key into the lock and turning clockwise into the locked
position.