Chapter 2: Oximeter Circuit Description
3304 Service Manual
2-11
voltage Test Points
Unless otherwise noted, all voltages are measured with respect to
ground at TP-1.
LOCATIOn
COnDITIOn
nOMInAL AnD RAnge
TP2
Oximeter on.
5 VDC, ± 0.25 VDC
ANA+5
Same as TP2.
Same as TP2.
FLTR+5
Same as TP2.
Same as TP2.
TP3
Oximeter on.
-4.8 VDC to -6.2 VDC
ANA-5
Same as TP3.
Same as TP3.
FLTR-5
Same as TP3.
Same as TP3.
V-REF
Oximeter on.
2.500 VDC
V-RAM
Oximeter on or off.
5VDC, ± 0.25 VDC.
U10-71
Oximeter on.
6.144 MHz, 50% duty cycle.
TP7
Referred to TP6.
4.7 VDC to 5.3 VDC
TP5
Oximeter on or off.
5.2 VDC to 7.2 VDC depending on the
state of the charge of the battery.
TP14
Oximeter is on.
3.65 VDC ± 0.1 VDC.
Waveform Test Points
The waveforms are measured with respect to ground at TP1.
•
The oscilloscope is triggered on TP16, falling edge.
•
The waveforms are measured with the finger probe attached to
•
the oximeter with finger or piece of paper in probe.