CADD-Prizm® VIP

Ambulatory Infusion Pump

Model 6100 and 6101

Patient Information

TPN 

Total Parenteral Nutrition

Delivery Mode

a

40-3665-01G PIG, Prizm TPN,VIP.indd   1

6/14/2011   10:50:21 AM

Summary of Contents for 6100

Page 1: ...x s CADD Prizm VIP Ambulatory Infusion Pump Model 6100 and 6101 Patient Information TPN Total Parenteral Nutrition Delivery Mode a ...

Page 2: ...er Pack or AC Adapter 9 The Main Screen 11 Description of the Infusion Profile 12 How to Use the HELP Key Å 13 Stopping and Restarting the Pump 15 Removing a Cassette 16 Attaching a Cassette 17 Priming the Tubing and Starting the Pump 20 Inserting the Tubing into the Air Detector 22 What if I drop or hit the pump 23 Alarms and Messages 24 ...

Page 3: ... parenteral nutrition solution or other fluids at a constant rate Delivery may taper up or gradually increase at the beginning of delivery Delivery may also taper down or gradually decrease at the end of delivery The pump stores programmed information and historical information that your clinician needs for your specific therapy Your clinician will instruct you on the proper use of this pump This ...

Page 4: ...egularly compare the volume remaining in the syringe to the pump s displayed values such as Reservoir Volume or Given to determine if under delivery is occurring and if necessary contact your clinician Do not use rechargeable NiCd or nickel metal hydride NiMH batteries Do not use carbon zinc heavy duty batteries They do not provide sufficient power for the pump to operate properly Always have new ...

Page 5: ...from an improperly attached cassette For detailed instructions and warning pertaining to CADD CADD Medication Cassette Reservoirs or CADD Administration Sets please refer to the instructions for use supplied with those products Attach the cassette the part of the CADD Medication Cassette Reservoir or CADD Administration Set that attaches to the pump properly An improperly attached or detached cass...

Page 6: ...ypad or enter the battery compartment Data In Out jack accessory jack or Air Detector Port area Do not clean the pump with acetone other plastic solvents or abrasive cleaners Use only Smiths Medical accessories as using other brands may adversely affect the operation of the pump Do not store the pump for prolonged periods with the battery installed Failure to push the Power Pack cord connector or ...

Page 7: ...servoir or CADD Administration Set that attaches to the pump Battery Compartment Green Light Blinks every 3 seconds when the pump is running and delivering fluid Amber Light Flashing pump is stopped or an alarm exists Steady pump is inoperable call clinician Display Shows information and messages After a short time the display turns itself off to save power Press any key to turn the display back o...

Page 8: ...ey explains what you see on the display Used by the clinician Advances you from one screen to the next and silences some alarms Í This key is not used in the TPN Delivery mode Used by the clinician Lets you answer yes to a question on the pump s display Î Lets you answer no to a question on the pump s display ...

Page 9: ...ot provide sufficient power for the pump to operate properly which could result in death or serious injury Always have new batteries available for replacement If power is lost non delivery of drug will occur and depending on the drug being administered could result in death or serious injury There is no pump alarm to alert you that the battery has not been properly installed or has become dislodge...

Page 10: ... press NOTE If you put the battery in backwards the display will remain blank Reinsert the battery making sure to match the and markings The power up sequence will start the pump will go through an electronic self test and the pump will beep at the end of the power up sequence All of the display indicators the software revision level and each parameter will appear briefly CAUTION Do not store the ...

Page 11: ...s a backup Otherwise you won t be able to start the pump Your Power Pack or AC Adapter may or may not have the connector shown 1 Open the Power cover on the side of your pump 2 Line up the red mark on the Power Pack cord connector with the red mark on the pump s Power jack 3 Push the connector forward until it stops Do not twist or turn the connector Pull lightly on the cord directly behind the fl...

Page 12: ...pter connector to the Power Pack input jack Or if you want to operate the pump on AC power you can attach the AC Adapter directly to the pump s Power jack AC ADAPTER POWER PACK Detaching the Power Pack 1 Grasp the connector 2 Pull the connector back using a straight steady motion DO NOT twist or turn the connector ...

Page 13: ... called the Main Screen and shows the following Battery Status TPN 6230X Low Battery RUNNING Res Vol 3000 0 ml Press NEXT to advance Reminder that the key lets you look at the pump s program Status of Infusion Period See the next page If the pump is stopped this will show STOPPED Status of Reservoir Volume ...

Page 14: ...he infusion profile is reset so delivery starts here At the end of the infusion profile you ll hear 9 beeps The KVO rate starts automatically RUNNING K KVO RUNNING Taper up RUNNING Plateau RUNNING Taper down Your clinician may or may not have programmed a taper up or a taper down If there is no taper up delivery will start at the plateau rate If there is no taper down you will still hear 9 beeps a...

Page 15: ...ppear The symbol in the lower right corner means there are more help screens Press Åagain to see the next help screen To page through all the help screens press Årepeatedly This will take you through all the help screens then back to the original screen you had questions about To return to the original screen you had questions about without going through all the help screens press This is the calc...

Page 16: ...lism An undetected upstream occlusion may result in under or non delivery of drug and depending on the drug being administered could result in death or serious injury Air embolism can result in death or serious injury 1 Press 2 When Start the pump appears press Starting pump will appear The pump will automatically review and display the preprogrammed settings RUNNING will appear on the main screen...

Page 17: ...VO delivery the Infusion Profile and the Reservoir Volume will automatically be reset This means that when you restart the pump it will start from the beginning of the Infusion Profile 1 Press 2 When Stop the Pump appears press You will see two screens if you stopped the pump during KVO delivery STOPPED will appear on the main screen and the amber light will blink Stop the Pump Press Y or N TPN 62...

Page 18: ...ng clamps and disconnect the tubing from your access device as instructed by your clinician 3 Use a coin to unlatch the cassette Insert the coin into the slot and turn clockwise until the latching button pops out NOTE If you cannot turn the latching button the cassette may be locked Contact your clinician 4 Remove the cassette from the pump 5 Discard the CADD Medication Cassette Reservoir or CADD ...

Page 19: ... in death or serious injury For detailed instructions and warnings pertaining to the CADD Medication Cassette Reservoir or CADD Administration Set please refer to the instructions for use supplied with those products CAUTION Frozen medication must be thawed at room temperature only Do not heat the CADD Medication Cassette Reservoir in a microwave oven as this may damage the medication the CADD Med...

Page 20: ...ssette Reservoir or CADD Administration Set that attaches to the pump properly An improperly attached or detached cassette could result in unregulated gravity infusion of medication from the fluid container or a reflux of blood which could result in death or serious injury 5 Gently twist and pull on the cassette to make sure it is firmly attached 6 A message will appear on the display showing the ...

Page 21: ...oir Volume to may appear Press to reset the Reservoir Volume to the originally programmed value This will also reset the Infusion Profile Reset Reservoir Volume to 3000 0 ml Press Y or N Infusion Profile has been reset ...

Page 22: ...catheter This could result in over delivery of medication or air embolism which could result in death or serious injury 3 Press and hold the key until the tubing is fully primed or until priming stops 4 This screen will appear if you release the key or if priming automatically stops If the tubing is not yet fully primed press the key and repeat step 3 If you are finished priming press the Îkey Pri...

Page 23: ...pen all clamps 6 Press to start the pump Starting pump will appear The pump will automatically review and display the programmed settings RUNNING will appear on the main screen the green light will blink and fluid delivery will begin as programmed by your clinician WARNING Ensure that the entire fluid path is free of all air bubbles before connecting to your catheter to prevent air embolism Air em...

Page 24: ...inician has instructed you to use the Air Detector open the Air Detector door and thread the tubing through the groove 2 Close the door making sure the tubing does not get pinched or kinked 3 Follow your clinician s instructions for connecting the tubing to your access site Open all clamps WARNING Ensure that the entire fluid path is free of all air bubbles before connecting to your catheter to pr...

Page 25: ...y twist and pull on the cassette to make sure it is still firmly attached Check the battery door to make sure it is still firmly attached If the cassette or battery door are loose or damaged do not use the pump Immediately stop the pump close the tubing clamp and contact your clinician WARNING If the pump is dropped or hit inspect the pump for damage Do not use a pump that is damaged or is not fun...

Page 26: ...Install a new 9 volt battery The pump will not run with a depleted 9 volt battery A good battery must be installed even when an external source of power is connected blank display or random characters A two tone alarm is sounding and the amber light stays on The 9 volt battery is depleted Install a new battery Power Pack Depleted Change Power Source The Power Pack is depleted Recharge the Power Pa...

Page 27: ... container and pump Press to stop the pump and silence the alarm for 2 minutes then remove the obstruction and press to restart the pump High Pressure Pump Stopped There may be a kink in the tubing or a clamp may be closed Unkink the tubing or open the clamp and the pump will resume delivery You may press to stop the pump and silence the alarm for 2 minutes After you remove the cause of high press...

Page 28: ...r Volume value is low indicating that the level of fluid in the fluid container is low Upstream Occlusion Press STOP to silence Model 6101 only Fluid is not flowing from the fluid container to the pump Check for a kink a closed clamp or an air bubble in the tubing between the fluid container and pump Press to stop the pump and silence the alarm for 2 minutes then remove the obstruction and press t...

Page 29: ...mp Check for a kink a closed clamp or an air bubble in the tubing between the fluid container and pump Press to stop the pump and silence the alarm then remove the obstruction and press to restart the pump Error Detected E error code There is a problem with the pump Close the tubing clamp and remove the pump from service Contact your clinician Alarms and Messages ...

Page 30: ...mpplacementduringbathing showering __________________ __________________________________________________________ Pumpplacementduringsleep _______________________________ __________________________________________________________ Storageofmedication _____________________________________ __________________________________________________________ Notes ...

Page 31: ......

Page 32: ... trademark is registered in the U S Patent and Trademark Office and certain other countries All other names and marks mentioned are the trademarks or service marks of their respective owners 1997 2011 Smiths Medical All rights reserved Smiths Medical ASD Inc St Paul MN 55112 USA 1 800 258 5361 USA 1 214 618 0218 www smiths medical com s ...