Appendix A
Pulse Oximeter for Sleep Screening Operation Manual
A-1
Description
The device will send out Trend Data to a PC through the sensor connector. To accomplish this,
the sensor must be disconnected from the device. Additional cables are required to connect
the device to a PC. The communication device will be dormant if the system is not in the Trend
Printing Mode.
Alternative PC Communication Setup (Not Using 1621 Software)
The following items will be needed:
Printer Adapter Cable (Catalog No. 3350)
DB-9 Modem Cable (Catalog No.3339)
NOTE! The DB9 PC Adapter Cable is an industry standard RS-232 null modem cable.
Power up the monitor.
Connect the printer communication cable labeled “BCI” to the monitor.
Connect the standard RS-232 null modem cable’s DB-9 connector to the mating connector of
the printer cable labeled “Printer”.
Connect the standard RS-232 null modem cable’s DB connector to the mating connector on
the PC.
NOTE! The DB-9 modem cable is an industry standard RS-232 null modem cable.
Figure A.1: PC Communication Setup
•
•
1.
2.
3.
4.
Appendix A: Data Format
%SpO2
B P M
x
F
Printer
Cable
DB-9 Null
Modem
Cable
DB-9 Printer
Connector