Chapter 7: Printer
7-4
Pulse Oximeter for Sleep Screening Operation Manual
What You’ll Need for Printing
You’ll need these items to print trend printouts:
Oximeter with installed printer.
Printer Paper: 1-inch wide thermal, Catalog No. 6012.
Trend Printouts
Collecting Trend Data
Whenever the monitor is on, it stores one SpO
2
and one pulse rate reading every 4
seconds. The stored readings are called
trend
data. The monitor remembers trend data for
up to 12 hours of run-time, depending upon battery life.
Trend data will be preserved for about one and a half minutes, without battery power,
allowing battery replacement without losing trend data.
Trend memory will contain data for only one patient.
Trends will be cleared by pressing and holding the
F
key for 6 to 8 seconds, at power-up.
A “clr” message will flash to indicate that trend clear is about to happen.
New data will not overwrite old data when the end of memory is reached.
NOTE! Turning the monitor OFF and back ON during Trend Collection will “freeze” trend
information and place the monitor in the Trend Printing Mode. The monitor must
have the trend data cleared and reset to begin Trend Collection again.
•
•
•
•
•
•
•