Chapter 9: Maintenance
Pulse Oximeter for Sleep Screening Operation Manual
9-1
Schedule of Maintenance
MAINTENANCE
HOW OFTEN
BY DOING THIS
Replace battery.
When Low Battery
Indicator is flashing or
12-15 hours of battery
use has elapsed.
Follow the instructions for installing the batteries.
Disinfect the reusable
sensor.
Before attaching the
sensor to the patient.
Follow the instructions for cleaning the reusable
sensor.
Disinfect the monitor.
When necessary.
Remove the batteries from the unit.
NOTE! Print tend data before cleaning.
Trend data will be lost if batteries are
disconnected more than one and a half
minutes.
Wipe the surfaces of the monitor with a soft,
clean cloth dampened in isopropyl alcohol.
Use only a cloth that is dampened, not wet.
CAUTION! Do not allow isopropyl alcohol
or any liquid to enter any of
the openings on the monitor.
Evidence that liquid has been
allowed to enter the monitor
voids the warranty.
1.
2.
Storage
Whenever possible, the monitor should be stored at room temperature in a dry environment.
If it is necessary to store the monitor for an extended period of time, the unit should be packed
in its original shipping container. Storing the monitor for a long period of time may degrade the
battery capacity. Batteries should be removed from the monitor before storing.
Storage specifications are as follows:
Temperature:
-40
o
C to +75
o
C (-40
o
F to +167
o
F)
Relative Humidity:
10% to 95% (noncondensing)
Chapter 9: Maintenance